Background: Catheter-directed thrombolysis (CDT) is an effective and safe endovascular method used in critical limb ischemia and many other thromboembolic events. Ultrasound-assisted catheter-directed thrombolysis (US-CDT) is an emerging technique considered to accelerate thrombolysis and therefore is supposed to improve outcome.

Purpose: To evaluate the efficacy of US-CDT in comparison to standard CDT in vitro.

Material And Methods: A total of 69 sets of human venous blood were evaluated, each comprising a tube just treated with CDT, a tube treated with US-CDT, and a control tube. All tubes were kept under physiological conditions. Except for the controls, in all tubes 5 mg of tissue-type plasminogen activator was administered over the predetermined treatment interval. Thrombus mass was weighted at the end of the lysis intervals at 6 h or 24 h, respectively.

Results: CDT led to a mean thrombus reduction of 32% and ultrasound-assisted lysis led to a mean thrombus reduction of 41% ( < 0.001 for both). Thrombus reduction was significantly higher after US-CDT compared to CDT ( = 0.001). The better efficacy of US-CDT was mostly already apparent at early phases during thrombolysis and did further mildly increase over time (r = 0.24;  = 0.047).

Conclusion: In vitro US-CDT is significantly superior to standard CDT; this effect is apparent at an early timepoint of lysis and slightly further increases over time.

Download full-text PDF

Source
http://dx.doi.org/10.1177/02841851211061440DOI Listing

Publication Analysis

Top Keywords

catheter-directed thrombolysis
16
tube treated
8
led thrombus
8
thrombus reduction
8
thrombolysis
5
efficacy ultrasound
4
ultrasound assisted
4
catheter-directed
4
assisted catheter-directed
4
thrombolysis compared
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!