Background: Cardiac implantable electronic devices (CIEDs) are widely used according to consensus guidelines in various patient categories. The longevity of CIED is a major determinant of the frequency with which patients require device replacement. Given the mismatch between the useful life of the devices and patient survival, device replacement is often needed. There is a great variability in the criteria used by manufacturers to determine the longevity of pacemakers (PM), implantable defibrillators (ICDs), and devices for cardiac resynchronization therapy (CRT). Thus, a fair comparison and an effective device evaluation is often difficult.

Methods: The objective of this paper is to provide standardized criteria based on typical clinical settings for estimating the longevity of single and dual chamber ICDs, cardiac resynchronization defibrillators (CRT-D), single and dual chamber PMs, and cardiac resynchronization PMs (CRT- P) to be used in health technology assessment for an appropriate comparison among different devices.

Results: The proposed parameters, if applied to the current marketed devices, provide longevity values in the range 5-17.2 years.

Conclusion: The values of longevity with the non-standardized criteria used by the manufacturers result in higher maximum values respect to the proposed standardized criteria for CRT-D, CRTD-MPP, CRT-P, and Dual-chamber PM.

Download full-text PDF

Source
http://dx.doi.org/10.1080/17434440.2021.2013199DOI Listing

Publication Analysis

Top Keywords

standardized criteria
12
cardiac resynchronization
12
cardiac implantable
8
implantable electronic
8
electronic devices
8
device replacement
8
criteria manufacturers
8
single dual
8
dual chamber
8
longevity
6

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!