Microfluidic photoreactor to treat neonatal jaundice.

Biomicrofluidics

School of Chemical, Biological, and Environmental Engineering, Oregon State University, Corvallis, Oregon 97331, USA.

Published: December 2021

While in most cases, jaundice can be effectively treated using phototherapy, severe cases require exchange transfusion, a relatively risky procedure in which the neonate's bilirubin-rich blood is replaced with donor blood. Here, we examine extracorporeal blood treatment in a microfluidic photoreactor as an alternative to exchange transfusion. This new treatment approach relies on the same principle as phototherapy but leverages microfluidics to speed up bilirubin removal. Our results demonstrate that high-intensity light at 470 nm can be used to rapidly reduce bilirubin levels without causing appreciable damage to DNA in blood cells. Light at 470 nm was more effective than light at 505 nm. Studies in Gunn rats show that photoreactor treatment for 4 h significantly reduces bilirubin levels, similar to the bilirubin reduction observed for exchange transfusion and on a similar time scale. Predictions for human neonates demonstrate that this new treatment approach is expected to exceed the performance of exchange transfusion using a low blood flow rate and priming volume, which will facilitate vascular access and improve safety.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8616629PMC
http://dx.doi.org/10.1063/5.0066073DOI Listing

Publication Analysis

Top Keywords

exchange transfusion
16
microfluidic photoreactor
8
treatment approach
8
light 470 nm
8
bilirubin levels
8
blood
5
photoreactor treat
4
treat neonatal
4
neonatal jaundice
4
jaundice cases
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!