Background: Data from literature shows that antigen tests are rapid and helpful tools for diagnosis of COVID-19.

Aim: This work aimed to evaluate the performances of the Elecsys SARS-CoV-2 Antigen test, in comparison to RT-qPCR, the gold standard.

Methods: A total of 110 swabs were tested; according to rRT-PCR, 76 were positive, and 34 were negative. The swabs were processed by Elecsys SARS CoV 2 Antigen assay (Roche Diagnostics GmbH, Mannheim, Germany), an electrochemiluminescence immunoassay (ECLIA).

Results: In a first evaluation, the overall sensitivity and specificity were 85% and 100%, respectively. It was noted that most of the discordant cases had cycle threshold (Ct) values > 28. Therefore, it was assumed a new measure to evaluate sensitivity and specificity, then samples with Ct values < 28 were selected. In this way, it was achieved a Ct < 28 sensitivity of 94%. The level of agreement between the two tests was 89. 1% with κ value of 0.77 for total data and 95.9% with κ value of 0.95 for samples with < 28 Ct. The antigen test performs well in the presence of high viral loads, whereas lower levels are missed.

Conclusions: The comparison data obtained in this study support that this method seems a proper approach for rapid screening of patients with high SARS-CoV-2 viral load; however, the rate of sensitivity is highly Ct-dependent.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8631556PMC
http://dx.doi.org/10.1007/s11845-021-02863-1DOI Listing

Publication Analysis

Top Keywords

sensitivity specificity
8
evaluation eclia
4
antigen
4
eclia antigen
4
antigen detection
4
detection tests
4
tests screening
4
screening methods
4
methods covid-19
4
covid-19 comparison
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!