Accurate radiation dosimetry is required for radiation protection in various environments. Therefore, dosemeters and dose-rate meters must be calibrated in standard radiation fields. The National Metrology Institute of Japan (NMIJ) expands the energy range of x-ray reference field measurement up to 450 kV using a cylindrical graphite-walled cavity ionization chamber. Departure from the condition of the Spencer-Attix cavity theory was evaluated by comparing the measurement results obtained using the cavity ionization and the free-air ionization chambers, which are used as the primary standard up to a tube voltage of 250 kV. The calibration coefficients found using the spherical ionization chamber were in good agreement with those obtained by the free-air ionization chamber within relative standard uncertainties ( = 1) for N-200 and N-250 x-ray fields. Consistent calibration coefficients were obtained in the energy range 300-450 kV.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1088/2057-1976/ac3e89 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!