A robust, affordable and portable light emitting diode-based diagnostic tools (POLS-NIRDx) using a polarization-sensitive (linear as well as circular polarization) technique were designed and developed to quantify the degree of linear polarization (DOLP), degree of circular polarization (DOCP). The study was performed on malignant (invasive ductal carcinoma) and adjacent normal ex-vivo biopsy tissues excised from N = 10 patients at the operating wavelengths of 850 and 940 nm. The average DOLP and DOCP values were lower for malignant than adjacent normal while operating at 850 and 940 nm. The highest accuracy was observed for DOLP (100%) and DOCP (80%) while operating at 850 nm, which reduced (80% for DOLP and 65% for DOCP) at 940 nm. This pilot study can be utilized as a differentiating factor to delineate malignant tissues from adjacent normal tissues.