Objective: Evaluating convenient, wearable multi-frequency impedance pneumography (IP)-based respiratory monitoring in ambulatory persons with novel electrode positioning.

Methods: A wearable multi-frequency IP system was utilized to estimate tidal volume (TV) and respiratory timings in 14 healthy subjects. A 5.1 cm × 5.1 cm tetrapolar electrode array, affixed to the sternum, and a conventional thoracic electrode configuration were employed to measure the respective IP signals, patch and thoracic IP. Data collected during static postures-sitting and supine-and activities-walking and stair-stepping-were evaluated against a simultaneously-obtained spirometer (SP) volume signal.

Results: Across all measurements, estimated TV obtained from the patch and thoracic IP maintained a Pearson correlation coefficient (r) of 0.93 ± 0.05 and 0.95 ± 0.05 to the ground truth TV, respectively, with an associated root-mean-square error (RMSE) of 0.177 L and 0.129 L, respectively. Average respiration rates (RRs) were extracted from 30-second segments with mean-absolute-percentage errors (MAPEs) of 0.93% and 0.74% for patch and thoracic IP, respectively. Likewise, average inspiratory and expiratory timings were identified with MAPEs less than 6% and 4.5% for patch and thoracic IP, respectively.

Conclusion: We demonstrated that patch IP performs comparably to traditional, cumbersome IP configurations. We also present for the first time, to the best of our knowledge, that IP can robustly estimate breath-by-breath TV and respiratory timings during ambulation.

Significance: This work represents a notable step towards pervasive wearable ambulatory respiratory monitoring via the fusion of a compact chest-worn form factor and multi-frequency IP that can be readily adapted for holistic cardiopulmonary monitoring.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9199959PMC
http://dx.doi.org/10.1109/TBME.2021.3130540DOI Listing

Publication Analysis

Top Keywords

patch thoracic
16
respiratory timings
12
tidal volume
8
volume respiratory
8
impedance pneumography
8
wearable multi-frequency
8
respiratory monitoring
8
respiratory
5
thoracic
5
patch
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!