AI Article Synopsis

  • - Spiral fMRI is an alternative to traditional echo-planar imaging, offering faster acquisition speeds and higher efficiency, making it suitable for applications that need detailed imaging like laminar fMRI.
  • - However, spiral fMRI faces challenges like blurring artifacts due to magnetic field imperfections; recent advancements in signal modeling and iterative reconstruction have addressed these issues.
  • - The study demonstrates the effectiveness of high-resolution spiral fMRI at 7 Tesla, achieving excellent image quality and activation maps, and showcases its versatility with combined readouts to enhance sensitivity further.

Article Abstract

Spiral fMRI has been put forward as a viable alternative to rectilinear echo-planar imaging, in particular due to its enhanced average k-space speed and thus high acquisition efficiency. This renders spirals attractive for contemporary fMRI applications that require high spatiotemporal resolution, such as laminar or columnar fMRI. However, in practice, spiral fMRI is typically hampered by its reduced robustness and ensuing blurring artifacts, which arise from imperfections in both static and dynamic magnetic fields. Recently, these limitations have been overcome by the concerted application of an expanded signal model that accounts for such field imperfections, and its inversion by iterative image reconstruction. In the challenging ultra-high field environment of 7 Tesla, where field inhomogeneity effects are aggravated, both multi-shot and single-shot 2D spiral imaging at sub-millimeter resolution was demonstrated with high depiction quality and anatomical congruency. In this work, we further these advances towards a time series application of spiral readouts, namely, single-shot spiral BOLD fMRI at 0.8 mm in-plane resolution. We demonstrate that high-resolution spiral fMRI at 7 T is not only feasible, but delivers both excellent image quality, BOLD sensitivity, and spatial specificity of the activation maps, with little artifactual blurring. Furthermore, we show the versatility of the approach with a combined in/out spiral readout at a more typical resolution (1.5 mm), where the high acquisition efficiency allows to acquire two images per shot for improved sensitivity by echo combination.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.neuroimage.2021.118738DOI Listing

Publication Analysis

Top Keywords

spiral fmri
16
high acquisition
8
acquisition efficiency
8
single-shot spiral
8
fmri
7
spiral
7
advances spiral
4
fmri high-resolution
4
high-resolution study
4
study single-shot
4

Similar Publications

Background: Postural tremor is a common clinical situation. Timely and accurate diagnosis is essential for effective treatment. However, clinicians often encounter difficulties distinguishing between essential tremor and other etiologies due to overlapping symptoms and atypical features.

View Article and Find Full Text PDF

Functional MRI (fMRI) is an important tool for non-invasive studies of brain function. Over the past decade, multi-echo fMRI methods that sample multiple echo times has become popular with potential to improve quantification. While these acquisitions are typically performed with Cartesian trajectories, non-Cartesian trajectories, in particular spiral acquisitions, hold promise for denser sampling of echo times.

View Article and Find Full Text PDF

Background: Contemporary 0.55T magnetic resonance imaging (MRI) is promising for fetal MRI, due to the larger bore, reduced safety concerns, lower acoustic noise, and improved fast imaging capability. In this work, we explore improved fetal cardiovascular magnetic resonance (CMR) without relying on any synchronizing devices, prospective, or retrospective gating, to determine the feasibility of real-time MRI evaluation of fetal cardiac function as well as cardiac and great vessel anatomies by using spiral balanced steady-state free precession (bSSFP) at 0.

View Article and Find Full Text PDF

Magnetic resonance acoustic radiation force imaging (MR-ARFI) is an exceptionally promising technique to non-invasively confirm targeting accuracy and estimate exposure of low-intensity transcranial focused ultrasound stimulation. MR-ARFI uses magnetic field motion encoding gradients to visualize the MR phase changes generated by microscopic displacements at the ultrasound focus. Implementing MR-ARFI in the human central nervous system has been hindered by 1) phase distortion caused by subject motion, and 2) insufficient signal-to-noise ratio at low (<1.

View Article and Find Full Text PDF
Article Synopsis
  • * Immediate surgery is crucial for type A AD (affecting the ascending aorta) to reduce mortality, while type B AD (descending aorta) is generally managed with medication unless complications arise.
  • * Case studies illustrate the severity of type A AD by showing fatal outcomes in two patients, while cases of type B had better results, emphasizing the need for increased awareness and improved healthcare policies for cardiovascular emergencies.
View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!