Long-term data of the new transcutaneous partially implantable bone conduction hearing system Osia®.

Eur Arch Otorhinolaryngol

Department of Oto-Rhino-Laryngology, Medical Center, University of Freiburg, Killianstr. 5, 79106, Freiburg, Germany.

Published: September 2022

Purpose: The new active transcutaneous partially implantable osseointegrated system Cochlear™ Osia® System is indicated in case of conductive or mixed hearing loss (CHL/MHL) with a maximum average bone conduction hearing loss of 55 dB, or in single-sided deafness (SSD). The implant directly stimulates the bone via a piezoelectric transducer and is directed by an external sound processor. We conducted a monocentric retrospective longitudinal within-subject clinical study at our tertiary academic referral center. The aim was to investigate long-term data (2017-2021) on audiological outcomes and hearing-related quality of life for the Osia system.

Methods: Between 2017 and 2020, 22 adults (18: CHL/MHL; 3: SSD) were implanted with the Osia100 implant; seven received bilateral implants. As of 10/2020, the sound processor was upgraded to Osia 2.

Results: Mean Osia system use by 04/2021 was 30.9 ± 8.6 months (range 17-40 months). Unaided bone conduction thresholds were unchanged postoperatively. One patient had to be explanted because of prolonged wound infection. Aided hearing thresholds were significantly lower compared to the unaided thresholds preoperatively, along with a marked increase in speech recognition in quiet. Speech processor upgrade resulted in a stable benefit. Patients with CHL/MHL and SSD showed a similar improvement in self-rated hearing performance revealed by SSQ, APHAB, and HUI questionnaires.

Conclusion: The Osia system is a safe, effective and sustainable option for treatment of conductive and mixed hearing loss or single-sided deafness.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9363475PMC
http://dx.doi.org/10.1007/s00405-021-07167-9DOI Listing

Publication Analysis

Top Keywords

bone conduction
12
hearing loss
12
long-term data
8
transcutaneous partially
8
partially implantable
8
conduction hearing
8
conductive mixed
8
mixed hearing
8
single-sided deafness
8
sound processor
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!