Objective: Autologous bone grafting for cranioplasty is associated with a high infection rate and bone absorption. Synthetic implant materials for cranioplasty have been developed. In this study, we evaluated the efficacy of titanium mesh-type patient-specific implants (PSIs) for patients with skull defects using the dice similarity coefficient (DSC), clinical outcomes, and artifacts caused by implants.

Methods: This retrospective study included 40 patients who underwent cranioplasty with a titanium mesh PSI at our institution. Based on preoperative and postoperative computed tomography scans, we calculated DSC and artifacts.

Results: The calculated DSC of 40 patients was 0.75, and the noise was 13.89% higher in the region of interest (ROI) near the implanted side (average, 7.64 hounsfield unit [HU]±2.62) than in the normal bone (average, 6.72 HU±2.35). However, the image signal-to-noise ratio did not significantly differ between the ROI near the implanted side (4.77±1.78) and normal bone (4.97±1.88). The patients showed no significant perioperative complications that required a secondary operation.

Conclusion: Titanium mesh-type PSIs for cranioplasty have excellent DSC values with lower artifacts and complication rates.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8558026PMC
http://dx.doi.org/10.13004/kjnt.2021.17.e25DOI Listing

Publication Analysis

Top Keywords

titanium mesh-type
12
mesh-type patient-specific
8
calculated dsc
8
roi implanted
8
implanted side
8
normal bone
8
cranioplasty
5
efficacy 3d-printed
4
titanium
4
3d-printed titanium
4

Similar Publications

Objective: Total en bloc spondylectomy (TES) is a curative surgical method for spinal tumors. After resecting the 3 spinal columns, reconstruction is of paramount importance. We present cases of mechanical failure and suggest strategies for salvage surgery.

View Article and Find Full Text PDF

Objective: Autologous bone grafting for cranioplasty is associated with a high infection rate and bone absorption. Synthetic implant materials for cranioplasty have been developed. In this study, we evaluated the efficacy of titanium mesh-type patient-specific implants (PSIs) for patients with skull defects using the dice similarity coefficient (DSC), clinical outcomes, and artifacts caused by implants.

View Article and Find Full Text PDF

Assessment of short-term outcome with TiO mesh in laparoscopic repair of large paraesophageal hiatal hernias.

BMC Surg

October 2019

Division of Esophagogastric, Bariatric and Minimally Invasive Surgery, Department of Surgery, Ramon y Cajal University Hospital, Crta. Colmenar Viejo Km 9,100, 28034, Madrid, Spain.

Background: Laparoscopic large para-oesophageal hiatal hernia (LPHH) repair using mesh reinforcement significantly reduces postoperative recurrence rates compared to conventional suture repair, especially within short follow-up times. However, the ideal strategy for repairing LPHH remains disputable because no clear guidelines are given regarding indications, mesh type, shape or position. The aim of this study was to survey our short-term results of LPHH management with a biosynthetic monofilament polypropylene mesh coated with titanium dioxide to enhance biocompatibility (TiOMesh™).

View Article and Find Full Text PDF

Objective: To evaluate the histological response to and changes in the biomechanical properties of titanized polypropylene lightweight mesh and conventional polypropylene mesh at 1 and 12 weeks following implantation in the sheep vagina.

Methods: We compared a titanized polypropylene lightweight mesh (TiLOOP Mesh) to a conventional polypropylene mesh (Gynemesh PS) in a sheep vagina model. Explants were harvested after 1 and 12 weeks (n = 6/mesh type/time point) for histological observation.

View Article and Find Full Text PDF

Surgical Correction of Diplopia in Orbital Fracture: Influence of Material and Design.

Ann Maxillofac Surg

January 2019

Director and Consultant, Oral and Craniomaxillofacial Surgeon, Balaji Dental and Craniofacial Hospital, Chennai, Tamil Nadu, India.

Purpose: The purpose of this study is to compare the association of diplopia after orbital fracture repair using titanium mesh and high-density polyethylene-coated titanium mesh.

Materials And Methods: Retrospective review of records of consecutive patients who presented for primary/secondary correction of orbital fracture attending author's institute was done. Patients attending the institution between January 2013 and December 2017 (5 years' period) fulfilling the inclusion and exclusion criteria were included for this study.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!