Talus implants can be utilized in cases of talus avascular necrosis and has been regarded as a promising treatment method. However, existing implants are made of stiff materials that directly oppose natural cartilage. The risk of long-term cartilage wear and bone fracture from the interaction between the cartilage and stiff implant surfaces has been documented in post-hemiarthroplasty of the hip, knee and ankle joints. The aim is to explore the effects of adding a layer of compliant material (polycarbonate-urethane; PCU) over a stiff material (cobalt chromium) in talus implants. To do so, we obtained initial ankle geometry from four cadaveric subjects in neutral standing to create the finite element models. We simulated seven models for each subject: three different types of talus implants, each coated with and without PCU, and a biological model. In total, we constructed 28 finite element models. By comparing the contact characteristics of the implant models with their respective biological model counterparts, our results showed that PCU coated implants have comparable contact area and contact pressure to the biological models, whereas stiff material implants without the PCU coating all have relatively higher contact pressure and smaller contact areas. These results confirmed that adding a layer of compliant material coating reduces the contact pressure and increases the contact area which in turn reduces the risk of cartilage wear and bone fracture. The results also suggest that there can be clinical benefits of adding a layer of compliant material coating on existing stiff material implants, and can provide valuable information towards the design of more biofidelic implants in the future.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jmbbm.2021.104936DOI Listing

Publication Analysis

Top Keywords

talus implants
12
adding layer
12
layer compliant
12
compliant material
12
stiff material
12
contact pressure
12
contact
8
contact characteristics
8
implants
8
cartilage wear
8

Similar Publications

Objective: To explore weight-bearing stability of Pilon fracture fixed by external fixator.

Methods: Six ankle bone models (right side) and 4 pairs (8 ankle cadaver specimens) were selected. Pilon fracture model was prepared by using the preset osteotomy line based on Ruedi Allgower Pilon fracture type.

View Article and Find Full Text PDF

[Navigated and minimally invasive screw osteosynthesis of a talus fracture].

Unfallchirurgie (Heidelb)

December 2024

Klinik für Unfallchirurgie, Technische Universität München, Klinikum rechts der Isar, Ismaninger Straße 22, 81675, München, Deutschland.

Objective Of Surgery: The aim of this surgery is to safeguard the multifragmentary and nondisplaced talus fracture (body and neck) against secondary dislocation in a navigated and minimally invasive manner using screw osteosynthesis.

Indications: Due to the young age of the patient in the presented case and the risk of a possible secondary dislocation, the decision was made in favor of surgical treatment.

Contraindications: Soft tissue swelling, wound infections and allergies to the osteosynthesis material.

View Article and Find Full Text PDF

Background: Nonprimary osteochondral lesions of the talus (OLT) pose a significant challenge in orthopaedics, with no definitive consensus on optimal surgical treatment.

Purpose: To consolidate the most recent evidence on operative treatments for nonprimary OLT by assessing patient-reported outcomes (PROs), postoperative complications, and clinical failures.

Study Design: Systematic review; Level of evidence, 4.

View Article and Find Full Text PDF

Introduction: Subtalar joint arthrodesis is primarily indicated for advanced osteoarthritis, hindfoot deformity, and/or instability. During the first 6-10 weeks after surgery, there is an intermediary structurally weaker state before complete bony fusion of the calcaneus and talus occurs. Loading of the foot can lead to mechanical stresses and relative movements in the former joint gap, which can impede the fusion process.

View Article and Find Full Text PDF

Total talus replacement (TTR) is a viable surgical option in the setting of talar avascular necrosis and collapse, as well as a revision option for failed total ankle arthroplasty with talar implant subsidence. The purpose of the present study was to compare the clinical and radiographic outcomes following isolated TTR and TTR combined with total ankle arthroplasty or hindfoot arthrodesis. Patients who underwent TTR were retrospectively reviewed, as a multicenter consecutive case series.

View Article and Find Full Text PDF