In this study, we developed an analytical method for simultaneous determination of 14 quinolones and 4 tetracyclines in livestock and fishery products using LC-MS/MS. The analytes were extracted from food samples with citrate buffer (containing EDTA)-methanol-acetonitrile (3 : 1 : 1, v/v/v) in the presence of n-hexane, and the extract was purified with an Oasis PRiME HLB cartridge column. It was suggested that this analytical method can also extract analytes from solid samples containing fat by using n-hexane. In addition, using methanol-acetonitrile (3 : 7, v/v) containing 0.1 vol% formic acid as an eluent from the cartridge column, the purification effect could be improved, while minimizing the impairment of the recovery rate. As a result of the validation using six types of food samples, trueness (accuracy) was 70.6%-113.8%, the RSD of repeatability was 9.0% or less, and the RSD of within-laboratory reproducibility was 15.5% or less. Using this approach, the standard values mentioned in the Japanese guideline were successfully met.