Purpose: To compare quantitative biliary measurements obtained with three different magnetic resonance cholangiopancreatography (MRCP) acquisition methods.

Methods: This retrospective study was IRB-approved. Patients with combinations of clinically indicated 3D FSE MRCP with sensitivity encoding (SENSE), 3D FSE SENSE MRCP with compressed sensing (CS-FSE; acceleration factor 8), and 3D gradient and spin-echo (GRASE) MRCP, acquired between October 2018 and March 2020, were included. The MRCP + Tuning Threshold algorithm (Perspectum Ltd., Oxford, UK) was used to segment 3D biliary models from MRCP data, with multiple metrics quantified from the models. Single measure, two-way, mixed-effects intra-class correlations, Bland-Altman analyses, and Wilcoxon signed-rank tests were used to compare quantitative measurements.

Results: From 160 MRCP datasets (25 3D FSE, 67 3D CS-FSE, 68 3D GRASE) in 69 patients, 48 datasets (7 [28%] 3D FSE, 14 [21%] 3D CS-FSE, 27 [40%] 3D GRASE) failed post-processing due to motion artifacts. The remaining 112 MRCP datasets (18 3D FSE, 53 3D CS-FSE, 41 3D GRASE) from 60 patients were included in the analysis. There was good to excellent agreement between 3D FSE and 3D CS-FSE MRCP for diameter of the left and right hepatic ducts, biliary volume, number and length of ducts, and total length of dilations (ICC: 0.83-0.93). The only metrics that exhibited good agreement between 3D FSE and 3D GRASE MRCP were biliary volume (ICC: 0.75) and total number of dilations (ICC: 0.77).

Conclusion: 3D CS-FSE MRCP produces comparable biliary diameter metrics and global duct quantification to 3D FSE MRCP at a significantly reduced acquisition time.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00261-021-03330-2DOI Listing

Publication Analysis

Top Keywords

fse cs-fse
12
mrcp
11
magnetic resonance
8
measurements three
8
compare quantitative
8
fse
8
fse mrcp
8
grase mrcp
8
mrcp datasets
8
datasets fse
8

Similar Publications

Purpose: To compare quantitative biliary measurements obtained with three different magnetic resonance cholangiopancreatography (MRCP) acquisition methods.

Methods: This retrospective study was IRB-approved. Patients with combinations of clinically indicated 3D FSE MRCP with sensitivity encoding (SENSE), 3D FSE SENSE MRCP with compressed sensing (CS-FSE; acceleration factor 8), and 3D gradient and spin-echo (GRASE) MRCP, acquired between October 2018 and March 2020, were included.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!