Objective: This paper was a anatomical radiographic study of distance between lumbar bi-cortical pedicle screws (BPSs) and anterior large vessels (ALVs) in patients with lumbar spondylolisthesis, and to provide clinical basis for evaluating the safety of bi-cortical pedicle screw implantation during lumbar spondylolisthesis.

Methods: Complete Computed tomography (CT) data of 104 patients with grade I lumbar spondylolisthesis (L4 52 and L5 52) and 107 non-spondylolisthesis patients (control group) were collected in this study. The distances between lumbar 4,5(L4,5) and sacrum 1(S1) BPSs and ALVs (abdominal aorta, inferior vena cava, left and right common iliac artery, internal and external iliac artery) were respectively measured at different transverse screw angles (TSAs) (L4:5°,10°; L5:10°,15°; S1:0°,5°,10°) and analyzed by SPSS (v25.0). There were three types of distances from the anterior vertebral cortex (AVC) to the ALVs (D): D N, D ≥ 0.50 cm, and D < 0.50 cm; these different distances represented non-contact, distant and close ALV respectively.

Results: We calculated the incidences of screw tip contacting large vessels at different TSAs and provided the appropriate angle of screw implantation. In non-spondylolisthesis group, in L4, the appropriate left TSA was 5°, and the incidence of the close ALV was 4.62%. In S1, the appropriate left TSA was 0° and the incidence of the close ALV was 22.4%, while the appropriate right TSA was 10° and the incidence of the close ALV was 17.8%. In L4 spondylolisthesis group, in L4, the appropriate left TSA was 5°, and the incidence of the close ALV was 3.8%. In L5 spondylolisthesis group, in S1, the appropriate left TSA was 0° and the incidence of the close ALV was 19.2%, while the appropriate right TSA was 10° and the incidence of the close ALV was 21.2%. The use of BPS was not appropriate on the right side of L4 or on the either side of L5 both in spondylolisthesis and control group. In patients with lumbar 4 spondylolisthesis, the incidences of screw tip contacting large vessels were less than the control group in both L4 and 5. In patients with lumbar 5 spondylolisthesis, the incidences of screw tip contacting large vessels were less than the control group in L5, while there were no significant difference in S1.

Conclusion: It is very important that considering the anatomical relationship between the AVC and the ALVs while planning BPSs. The use of BPS does not apply to every lumbar vertebra. In patients with lumbar spondylolisthesis and non-spondylolisthesis patients, the incidences of screw tip contacting large vessels are different.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8561981PMC
http://dx.doi.org/10.1186/s12891-021-04811-7DOI Listing

Publication Analysis

Top Keywords

bi-cortical pedicle
12
lumbar spondylolisthesis
12
study distance
8
lumbar bi-cortical
8
pedicle screws
8
anterior large
8
large vessels
8
patients lumbar
8
iliac artery
8
lumbar
7

Similar Publications

Objective: This paper was a anatomical radiographic study of distance between lumbar bi-cortical pedicle screws (BPSs) and anterior large vessels (ALVs) in patients with lumbar spondylolisthesis, and to provide clinical basis for evaluating the safety of bi-cortical pedicle screw implantation during lumbar spondylolisthesis.

Methods: Complete Computed tomography (CT) data of 104 patients with grade I lumbar spondylolisthesis (L4 52 and L5 52) and 107 non-spondylolisthesis patients (control group) were collected in this study. The distances between lumbar 4,5(L4,5) and sacrum 1(S1) BPSs and ALVs (abdominal aorta, inferior vena cava, left and right common iliac artery, internal and external iliac artery) were respectively measured at different transverse screw angles (TSAs) (L4:5°,10°; L5:10°,15°; S1:0°,5°,10°) and analyzed by SPSS (v25.

View Article and Find Full Text PDF

Background And Objective: The technique of tri-cortical pedicle screw (TCPS) has been used to improve the anchoring strength in the sacral vertebrae. However, no studies have reported their application in the thoracic vertebrae. Our research is aimed to assess the stability and strength of the TCPS in thoracic vertebrae under osteoporotic condition by three dimensional (3D) finite element method on the basis of medical image reconstruction using computed tomography (CT), and verifying its effectiveness in clinical application.

View Article and Find Full Text PDF

Pedicle Screw Fixation Study in Immature Porcine Spines to Improve Pullout Resistance during Animal Testing.

PLoS One

June 2016

Aix-Marseille Université, CNRS, ISM UMR 7287, 13288, Marseille cedex 09, France; APHM, Hôpital Sainte-Marguerite, Institute for Locomotion, 13009, Marseille, France.

The porcine model is frequently used during development and validation of new spinal devices, because of its likeness to the human spine. These spinal devices are frequently composed of pedicle screws with a reputation for stable fixation but which can suffer pullouts during preclinical implantation on young animals, leading to high morbidity. With a view to identifying the best choices to optimize pedicle screw fixation in the porcine model, this study evaluates ex vivo the impact of weight (age) of the animal, the level of the vertebrae (lumbar or thoracic) and the type of screw anchorage (mono- or bi-cortical) on pedicle screw pullouts.

View Article and Find Full Text PDF

Lateral mass (LM) screws are commonly used in posterior instrumentation of the cervical spine because of their perceived safety over pedicle screws. A possible complication of cervical LM screw placement is vertebral artery injury or impingement. Several screw trajectories have been described to overcome the risks of neurovascular injury; however, each of these techniques relies on the surgeon's visual estimation of the trajectory angle.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!