Simulating dose reduction for myocardial perfusion SPECT using a Poisson resampling method.

Nucl Med Mol Imaging

Departments of Nuclear Medicine and Molecular Imaging, Ajou University School of Medicine, 164, World Cup-ro, Yeongtong-gu, Suwon, Kyunggi-do Republic of Korea 16499.

Published: October 2021

Purpose: The purpose of this study was to determine the lowest Tl-201 dose that does not reduce the image quality of myocardial perfusion SPECT (MPS) by Poisson resampling simulation.

Methods: One hundred and twelve consecutive MPS data from patients with suspected or known coronary artery disease were collected retrospectively. Stress and rest MPS data were resampled using the Poisson method with 33%, 50%, 67%, and 100% count settings. Two nuclear medicine physicians assessed the image quality of reconstructed data visually by giving grades from - 2 to + 2. The summed stress score (SSS), summed rest score (SRS), and summed difference score (SDS) were obtained on the workstation. Image quality grades and semi-quantitative scores were then compared among these resampled images.

Results: The proportions of "adequate" image quality were 0.48, 0.75, 0.92, and 0.96 for the groups of images with 33%, 50%, 67%, and 100% data, respectively. The quality of the resampled images was significantly degraded at 50% and 33% count settings, while the image quality was not different between 67 and 100% count settings. We also found that high body mass index further decreased image quality at 33% count setting. Among the semi-quantitative parameters, SSS and SRS showed a tendency to increase with a decline in count.

Conclusion: Based on the simulation results, Tl-201 dose for MPS can be reduced to 74 MBq without significant loss of image quality. However, the SSS and SRS can be changed significantly, and it needs to be further verified under the different conditions.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8517053PMC
http://dx.doi.org/10.1007/s13139-021-00710-wDOI Listing

Publication Analysis

Top Keywords

image quality
28
count settings
12
myocardial perfusion
8
perfusion spect
8
poisson resampling
8
tl-201 dose
8
quality
8
mps data
8
33% 50%
8
50% 67%
8

Similar Publications

Objective.: There is a growing consensus that interpersonal processes are key to understanding psychotherapy. How might that be reflected in the brain? Recent research proposes that inter-brain synchrony is a crucial neural component of interpersonal interaction.

View Article and Find Full Text PDF

Background And Purpose:  Treatment of idiopathic scoliosis in childhood aims to prevent curve progression. It is generally accepted that curves > 50° have the highest risk of progression, but less well described is what happens with mild to moderate curves. The aim of this study was to assess long-term curve progression and health-related quality of life (HRQoL) and compare thoracic and thoracolumbar/lumbar (TL/L) curves.

View Article and Find Full Text PDF

Objectives: The aim of this technical report was to assess whether the "Radiological Report" tool within the Artificial Intelligence (AI) software Diagnocat can achieve a satisfactory level of performance comparable to that of experienced dentomaxillofacial radiologists in interpreting cone-beam CT scans.

Methods: Ten cone-beam CT scans were carefully selected and analyzed using the AI tool, and they were also evaluated by two dentomaxillofacial radiologists. Observations related to tooth numeration, alterations in dental crowns, roots, and periodontal tissues were documented and subsequently compared to the AI findings.

View Article and Find Full Text PDF

Objectives: Due to the increasing use of cone-beam computed tomography (CBCT) in dentistry and considering the effects of radiation on radiosensitive organs, the aim of this study was to investigate the effect of shielding on absorbed dose of eyes, thyroid and breasts in scans conducted with different parameters using two different fields of view (FOV).

Methods: Dose measurements were calculated on a tissue-equivalent female phantom by repeating each scanning parameter three times and placing at least two thermoluminescent dosimeters (TLD) on each organ, with the averages then taken. The same CBCT scans were performed in two different FOV with shielding including thyroid collar, radiation safety glasses and lead apron and without shielding.

View Article and Find Full Text PDF

On the Effect of the Patient Table on Attenuation in Myocardial Perfusion Imaging SPECT.

EJNMMI Phys

January 2025

Institute of Radiology, Nuclear Medicine and Molecular Imaging, Heart and Diabetes Center North Rhine-Westphalia, University Hospital (Ruhr University Bochum), Medical Faculty OWL (Bielefeld University), Bad Oeynhausen, Germany.

Background: The topic of the effect of the patient table on attenuation in myocardial perfusion imaging (MPI) SPECT is gaining new relevance due to deep learning methods. Existing studies on this effect are old, rare and only consider phantom measurements, not patient studies. This study investigates the effect of the patient table on attenuation based on the difference between reconstructions of phantom scans and polar maps of patient studies.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!