Background: Occlusions of intravenous (IV) tubing can prevent vital and time-critical medication or solutions from being delivered into the bloodstream of patients receiving IV therapy. At low flow rates (≤ 1 ml/h) the alarm delay (time to an alert to the user) can be up to 2 h using conventional pressure threshold algorithms. In order to reduce alarm delays we developed and evaluated the performance of two new real-time occlusion detection algorithms and one co-occlusion detector that determines the correlation in trends in pressure changes for multiple pumps.
Methods: Bench-tested experimental runs were recorded in triplicate at rates of 1, 2, 4, 8, 16, and 32 ml/h. Each run consisted of 10 min of non-occluded infusion followed by a period of occluded infusion of 10 min or until a conventional occlusion alarm at 400 mmHg occurred. The first algorithm based on binary logistic regression attempts to detect occlusions based on the pump's administration rate Q(t) and pressure sensor readings P(t). The second algorithm continuously monitored whether the actual variation in the pressure exceeded a threshold of 2 standard deviations (SD) above the baseline pressure. When a pump detected an occlusion using the SD algorithm, a third algorithm correlated the pressures of multiple pumps to detect the presence of a shared occlusion. The algorithms were evaluated using 6 bench-tested baseline single-pump occlusion scenarios, 9 single-pump validation scenarios and 7 multi-pump co-occlusion scenarios (i.e. with flow rates of 1 + 1, 1 + 2, 1 + 4, 1 + 8, 1 + 16, and 1 + 32 ml/h respectively). Alarm delay was the primary performance measure.
Results: In the baseline single-pump occlusion scenarios, the overall mean ± SD alarm delay of the regression and SD algorithms were 1.8 ± 0.8 min and 0.4 ± 0.2 min, respectively. Compared to the delay of the conventional alarm this corresponds to a mean time reduction of 76% (P = 0.003) and 95% (P = 0.001), respectively. In the validation scenarios the overall mean ± SD alarm delay of the regression and SD algorithms were respectively 1.8 ± 1.6 min and 0.3 ± 0.2 min, corresponding to a mean time reduction of 77% and 95%. In the multi-pump scenarios a correlation > 0.8 between multiple pump pressures after initial occlusion detection by the SD algorithm had a mean ± SD alarm delay of 0.4 ± 0.2 min. In 2 out of the 9 validation scenarios an occlusion was not detected by the regression algorithm before a conventional occlusion alarm occurred. Otherwise no occlusions were missed.
Conclusions: In single pumps, both the regression and SD algorithm considerably reduced alarm delay compared to conventional pressure limit-based detection. The SD algorithm appeared to be more robust than the regression algorithm. For multiple pumps the correlation algorithm reliably detected co-occlusions. The latter may be used to localize the segment of tubing in which the occlusion occurs. Trial registration Not applicable.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8555317 | PMC |
http://dx.doi.org/10.1186/s12911-021-01668-7 | DOI Listing |
Microsyst Nanoeng
January 2025
Key Laboratory of Instrumentation Science and Dynamic Measurement Ministry of Education, North University of China, 030051, Taiyuan, China.
The alarming prevalence and mortality rates associated with cardiovascular diseases have emphasized the urgency for innovative detection solutions. Traditional methods, often costly, bulky, and prone to subjectivity, fall short of meeting the need for daily monitoring. Digital and portable wearable monitoring devices have emerged as a promising research frontier.
View Article and Find Full Text PDFBMC Public Health
January 2025
Department of Statistics, Faculty of Applied Sciences, Durban University of Technology, Durban, South Africa.
Background: Termination of pregnancy continues to be one of the major public health problems. The prevalence and determinants associated with pregnancy termination among women from low-middle income countries such as Tanzania have previously not been thoroughly investigated. Therefore, this study aims to explore the prevalence and determinants associated with pregnancy termination among Tanzanian women.
View Article and Find Full Text PDFJ Clin Med
December 2024
Clinical Department No. 10, "Carol Davila" University of Medicine and Pharmacy, 050474 Bucharest, Romania.
Controlling pain after major orthopedic surgery may be challenging, and it is related to delayed recovery, the development of chronic pain, and analgesic dependence. It is well known that effective postoperative pain control can reduce hospital stays by ensuring a more rapid rehabilitation,thereby decreasing the overall costs. Despite the development of analgesics, the use of opioids and their derivates remains the cornerstone of treatment for patients with acute moderate-to-severe pain in association with general or regional anesthesia.
View Article and Find Full Text PDFTrans R Soc Trop Med Hyg
January 2025
Department of Medicine, All India Institute of Medical Sciences, Jodhpur 342005, India.
Background: Snakebite is a neglected tropical disease that causes significant morbidity and mortality in India. In this study, we describe the clinical characteristics and outcomes of Echis carinatus sochureki envenoming from Western Rajasthan. We document the clinical ineffectiveness of the currently available Indian polyvalent antivenom in managing E.
View Article and Find Full Text PDFWorld J Orthop
December 2024
Law, Gujarat National Law University, Ahmedabad 382426, Gujarat, India.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!