Purpose: This is a multi-institutional report on inter-observer and inter-instrument variation in the calibration of the absorbed dose rate for a planar P beta emitting brachytherapy source. Measurement accuracy is essential since the dose profile is steep and the source is used for the treatment of tumors that are located in close proximity to healthy nervous system structures.

Methods And Materials: An RIC-100 P source was calibrated by three institutions using their own equipment and following their standard procedures. The first institution calibrated the source with an electron diode and EBT3 film. The second institution used an electron diode. The third institution used HD810 film. Additionally, each institution was asked to calibrate the source using an electron diode and procedure that was shared among all institutions and shipped along with the radiation source. The dose rate was reported in units of cGy*min at a water equivalent depth of 1 mm.

Results: Close agreement was observed in the measurements from different users and equipment. The variation across all diode detectors and institutions had a standard deviation of 1.8% and maximum difference of 4.6%. The observed variation among two different diode systems used within the same institution had a mean difference of 1.6% and a maximum variation of 1.8%. The variations among film and diode systems used within the same institution had a mean difference of 2.9% and a maximum variation of 4.3% CONCLUSIONS: The absorbed dose rate measurement protocol of the planar beta-emitting P source permits consistent dosimetry across three institutions and five different electron diode and radiochromic film systems. The methodologies presented herein should enable measurement consistency among other clinical users, which will help ensure high quality patient treatments and outcomes analysis.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8810704PMC
http://dx.doi.org/10.1016/j.brachy.2021.08.001DOI Listing

Publication Analysis

Top Keywords

dose rate
16
electron diode
16
absorbed dose
12
planar beta
8
beta emitting
8
emitting brachytherapy
8
source
8
brachytherapy source
8
three institutions
8
source electron
8

Similar Publications

Drug nanocrystal engineering is an attractive pharmaceutical approach to enhancing the oral bioavailability of poorly soluble drugs. The mechanism of drug nanocrystal stabilization, however, is unclear. Here we developed andrographolide nanocrystals (AG-NCs) with various nonionic surfactants (Pluronic-F127, TPGS, or Brij-S20).

View Article and Find Full Text PDF

Use of antibiotics in patients with chronic kidney disease: evidence from the real world.

Expert Opin Drug Saf

December 2024

Grupo de Investigación en Farmacoepidemiología y Farmacovigilancia, Universidad Tecnológica de Pereira-Audifarma S.A, Pereira, Risaralda, Colombia.

Background: Most antibiotics require dose adjustments in patients with chronic kidney disease (CKD) to avoid accumulation and toxicity. The aim was to characterize the use of systemic antibiotics in a group of patients with CKD and to adjust the dose according to the glomerular filtration rate (GFR).

Research Design And Methods: Observational study of patients with a diagnosis of CKD who received antibiotics between January-2021 July 2022.

View Article and Find Full Text PDF

Prediction of lacosamide concentrations to support dose optimization during pregnancy.

Epilepsia

December 2024

Department of Experimental and Clinical Pharmacology, College of Pharmacy, University of Minnesota, Minneapolis, Minnesota, USA.

Objective: We aimed to quantify and predict lacosamide exposure during pregnancy by developing a pregnancy physiologically-based pharmacokinetic model, allowing the prediction of potential dose increases to support maintaining a patient's preconception lacosamide concentrations.

Methods: Models for nonpregnant adults and pregnant female patients were constructed using physiochemical and pharmacological parameters identified from literature review. Evaluation of plasma concentration data from human males was digitized from the literature.

View Article and Find Full Text PDF

Aims: This phase I trial assessed the safety and potential efficacy of monthly 3 dose intravenous infusion of allogeneic bone marrow-derived clonal mesenchymal stromal cells (BM-cMSCs) in refractory rheumatoid arthritis (RA) patients over 24 weeks.

Patients & Methods: Six patients with refractory RA received BM-cMSC infusions at one-month intervals over a 24-week period. Safety outcomes included adverse events (AEs) and serious adverse events (SAEs).

View Article and Find Full Text PDF

Treatment and Follow-up of Non-stress Adrenal Insufficiency.

J Clin Res Pediatr Endocrinol

December 2024

Department of Pediatric Endocrinology and Diabetes, Derince Research and Training Hospital, Kocaeli, Turkey.

Adrenal insufficiency (AI) is defined as the inability of the adrenal cortex to produce adequate amounts of glucocorticoids and/or mineralocorticoids. As these hormones have important roles in water-salt balance and energy homeostasis, AI is a serious and potentially life-threatening condition. Glucocorticoid replacement therapy is vital in all cases of AI.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!