Attainable levels of signal-to-background ratio (SBR) in Raman spectroscopy of biological samples is limited by the presence of endogenous fluorophores. It is customary to remove the ubiquitous fluorescence background using postacquisition data processing. However, new approaches are needed to reduce background contributions and maximize the fraction of the sensor dynamical range occupied by Raman photons. Time-resolved detection using pulsed lasers and time-gated measurements can be used to address the signal-to-background problem in biological samples by limiting light detection to nonresonant interaction phenomena with relaxation time scales occurring on sub-nanosecond time scales, thereby excluding contributions from resonant phenomena such as fluorescence. A time-gated Fourier-transform spectrometer was assembled using a commercially available interferometer, a single channel single-photon avalanche diode and time tagging electronics. A time gate of 300 ps increased the signal-to-background-ratio of the 1440 cm Raman band from 36% to 69% in an olive oil sample hereby demonstrating the potential of this approach for autofluorescence suppression.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1002/jbio.202100188 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!