Experimental study of occlusion break surge volume in 3 different phacoemulsification systems.

J Cataract Refract Surg

From the Stein Eye Institute and Department of Ophthalmology, David Geffen School of Medicine at UCLA, Los Angeles, California (Miller); Alcon Research LLC, Lake Forest, California (Dyk, Yalamanchili).

Published: November 2021

Purpose: To evaluate surge volume after occlusion breaks in 3 different phacoemulsification systems.

Setting: Alcon Research LLC, Lake Forest, California.

Design: Experimental study.

Methods: A mechanical spring eye model was used to test the Centurion with Active Sentry, WhiteStar Signature Pro, and Stellaris PC. Raw oscilloscope data were converted to volumetric and pressure measurements. Fitted average surge traces were generated for each test condition and used to develop an interpolation algorithm to predict transient occlusion break surge events. Two-dimensional heat maps were generated.

Results: During occlusion break surge, the Centurion with Active Sentry had smaller aqueous volume losses than the WhiteStar or Stellaris PC. Maximum surge volumes within the mechanical spring eye model displacement limit were 74.7 μL, 157.5 μL, and 151.7 μL using Centurion with Active Sentry, WhiteStar, and Stellaris PC, respectively. In the aphakic state, heat maps showed that Centurion with Active Sentry had less than 20% aqueous volume loss across all vacuum limits and target intraocular pressure; WhiteStar and Stellaris PC systems had up to 35% and 50% aqueous volume losses, respectively, at the higher vacuum limits. In the phakic state, Centurion with Active Sentry had up to 30% aqueous volume loss and WhiteStar and Stellaris PC systems had up to 50% aqueous volume losses. In addition, predicted transient traces demonstrated that Centurion with Active Sentry had the lowest percentage simulated aqueous volume loss compared with WhiteStar or Stellaris PC.

Conclusions: Centurion with Active Sentry had lower aqueous volume losses after occlusion break than WhiteStar and Stellaris PC systems at all surgical settings.

Download full-text PDF

Source
http://dx.doi.org/10.1097/j.jcrs.0000000000000651DOI Listing

Publication Analysis

Top Keywords

centurion active
28
active sentry
28
aqueous volume
28
whitestar stellaris
24
occlusion break
16
volume losses
16
break surge
12
volume loss
12
stellaris systems
12
volume
9

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!