Background: Humidification of inspiratory gases is mandatory in all mechanically ventilated patients in ICUs, either with heated humidifiers (HHs) or with heat and moisture exchangers (HMEs). In patients with COVID-19, the choice of the humidification device may have relevant impact on patients' management as demonstrated in recent studies. We reported data from 2 ICUs using either HME or HH.
Methods: Data from patients with COVID-19 requiring invasive mechanical ventilation during the first wave in 2 ICUs in Québec City were reviewed. In one ICU, HMEs were used, whereas heated-wire HHs were used in the other ICU. We compared ventilator settings and arterial blood gases at day one after adjustment of ventilator settings. Episodes of endotracheal tube occlusions (ETOs) or subocclusions and a strategy to limit the risk of under-humidification were reported. On a bench test, we measured humidity with psychrometry with HH at different ambient temperature and evaluated the relation with heater plate temperature.
Results: We reported data from 20 subjects positive for SARS-Cov-2, including 6 in the ICU using HME and 14 in the ICU using HH. In the HME group, P was higher (48 vs 42 mm Hg) despite higher minute ventilation (171 vs 145 mL/kg/min predicted body weight [PBW]). We also reported 3 ETOs occurring in the ICU using HH. The hygrometric bench study reported a strong correlation between heater plate temperatures of the HH and humidity delivered. After implementation of measures to avoid under-humidification, including heater plate temperature monitoring, no more ETOs occurred.
Conclusions: The choice of the humidification device used in subjects with COVID-19 had a relevant impact on ventilation efficiency (increased CO removal with lower dead space) and on complications related to low humidity, including ETOs that may be present with heated-wire HHs when used with high ambient temperatures.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.4187/respcare.09314 | DOI Listing |
Sci Rep
December 2024
School of Energy Science and Engineering, Central South University, Changsha, 410083, China.
A three-dimensional numerical model of the vacuum sintering furnace was established, combined with the custom program of temperature-voltage feedback regulation. Through simulationand experimental validation, the heating and holding stage as well as the thermal hysteresis phenomenon of the furnace were analyzed, a dimensionless quantity of hysteresis temperature difference was proposed and calculated, the distribution of the electric field and temperature uniformity of the furnace were discussed in detail, while the structural improvement approach was proposed based on simulation. The results show that: during the heating process, the maximum of thermal hysteresis temperature difference between the graphite cylinder and the heating tube is 0.
View Article and Find Full Text PDFHardwareX
September 2024
Department of Electrical & Computer Engineering, Western University, 1151 Richmond St., London, Ontario N6A 5B9, Canada.
To produce samples for both material testing and molded sheets/parts, this article details an open-source scientific cold and hot press design. It consists of two independent and modular upper and lower plate (929 cm) assemblies each containing four 125 W insulated steel strip heaters. The steel housing for these heaters is entirely modular and designed for ease of manufacture, assembly, and customization.
View Article and Find Full Text PDFFront Microbiol
October 2024
Department of Medical and Surgical Sciences, Alma Mater Studiorum - University of Bologna, Bologna, Italy.
Sci Rep
November 2024
Department of Mechanical Engineering, Mettu University, Metu, Ethiopia.
Four distinct neural models were used to evaluate the efficiency of a V-trough solar water heater (VTSWH) equipped with square-cut twisted tape (SCTT) and V-cut twisted tape (VCTT) at two different twist ratios, 3 and 5. The objective of this study was the use of ANFIS (Adaptive Neuro-Fuzzy Inference System), G.L.
View Article and Find Full Text PDFEnviron Sci Pollut Res Int
November 2024
Department of Energy, School of Engineering, Tezpur University, Tezpur, India.
Heat storage materials improve the utility of solar air heaters (SAHs) after sunset. This study investigates an improved solar air heater (SAH) performance with baffles and waste mild steel chips as sensible heat storage (SHS) materials. Comparative experimental natural convection heat transfer studies were performed with four different improved air heater setups under similar solar radiation conditions.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!