Objectives: When using indirect ion-selective electrode (ISE) methods, hypertriglyceridemia leads to pseudohyponatremia due to water displacement artifacts. Multiple strategies exist to minimize this interference. Our objective was to create a patient-friendly one-tube-fits-all testing setup without compromising the method robustness.
Methods: Four strategies were evaluated in a single patient with hypertriglyceridemia. Additionally, the interchangeability between the Cobas 8000 and ABL Flex was evaluated on samples (n = 2274) with different total protein (TP) concentrations. Finally, a proof-of-concept (n = 40) was performed by re-measuring the routine sample with the ABL90 Flex.
Results: ABL90 flex results and calculated sodium did not suffer from the presence of high triglyceride levels. We did not observe any significant differences between the three groups (P > 0.05) of sample types (arterial vs. venous plasma vs. venous whole blood after mixing up) nor for the analysers (Roche vs. ABL90 Flex). Passing-Bablok and Bland-Altman tests revealed interchangeability.
Conclusion: In future cases of hypertriglyceridemia, 1500 mg/dL will be used as a preliminary threshold for reliable sodium determination. Routine Li-heparin samples can be used for accurate sodium determination without any need for extra arterial or venous blood gas tubes, offering a patient-friendly test setup for similar cases.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1080/17843286.2021.1992583 | DOI Listing |
Clin Biochem
January 2025
Pathology and Laboratory Medicine Program, Health Sciences Centre, St. John's, Newfoundland and Labrador, Canada; Memorial University of Newfoundland, Health Sciences Centre, St. John's, Newfoundland and Labrador, Canada. Electronic address:
Purpose: Rapid determination of cerebrospinal fluid (CSF) glucose and lactate is required by emergency rooms and intensive care units. Long turnaround time (TAT) on test results negatively impacts timely diagnosis and treatment of neurological infections like meningitis.
Methods: The CSF glucose and lactate assays were evaluated on a blood gas analyzer, Radiometer ABL90 Flex Plus.
Clin Chem Lab Med
December 2024
Department of Clinical Chemistry, 162073 Sestre Milosrdnice University Hospital Center, Zagreb, Croatia.
Objectives: Autoverification increases the efficiency of laboratories. Laboratories accredited according to ISO 15189:2022 need to validate their processes, including autoverification, and assess the associated risks to patient safety. The aim of this study was to propose a systematic verification algorithm for autoverification and to assess its potential risks.
View Article and Find Full Text PDFMedicina (B Aires)
December 2023
Asociación Bioquímica Argentina (ABA), Buenos Aires, Argentina.
Introduction: At the beginning of the SARS-CoV-2 pandemic, acute respiratory failure has been the most important cause of hospitalization in patients with COVID-19, being more severe in patients with comorbidities and risk factors. In these scenarios hypoxemia has been associated with increased mortality. Our objective was to identify parameters obtained from arterial blood gases (ABG) associated with mortality in patients with COVID-19 at hospital admission.
View Article and Find Full Text PDFScand J Trauma Resusc Emerg Med
December 2023
The Prehospital Research Unit, Region of Southern Denmark, Odense University Hospital, Odense, Denmark.
Background: The prehospital patients treated by ambulances and mobile emergency care units (MECU) are potentially critically ill or injured. Knowing the risks of serious outcomes in these patients is important for guiding their treatment. Some settings allow for prehospital arterial blood gas analyses.
View Article and Find Full Text PDFVet Clin Pathol
December 2023
Department of Clinical Chemistry, Karolinska University Hospital, Stockholm, Sweden.
Venous blood gases were analyzed with ABL90 FLEX in two cats with Heinz bodies in approximately 60% of the erythrocytes. The instrument demonstrated an inability to correctly report standard bicarbonate (stHCO ), hematocrits, and hemoglobin concentrations by indicating an OXI spectrum mismatch alarm (ie, the spectrum of measured hemoglobin forms differed from the spectrum of calculated forms). Actual bicarbonate (aHCO ) did not indicate any errors.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!