The Study of Field Equivalence Determined by the Modeled Percentage Depth Dose in Electron Beam Radiation Therapy.

Biomed Res Int

Heavy Ion Center of Wuwei Cancer Hospital, Gansu Wuwei Academy of Medical Sciences, Gansu Wuwei Tumor Hospital, Wuwei City Gansu Province, China.

Published: February 2022

Introduction: This study presents an empirical method to model the curve of electron beam percent depth dose (PDD) by using the primary-tail function in electron beam radiation therapy. The modeling parameters and can be used to predict the minimal side length when the field size is reduced below that required for lateral scatter equilibrium (LSE) in electron radiation therapy.

Methods And Materials: The electrons' PDD curves were modeled by the primary-tail function in this study. The primary function included the exponential function and the main parameters of and , while the tail function was composed of a sigmoid function with the main parameter of . The PDD of five electron energies was modeled by the primary and tail function by adjusting the parameters of , , and . The and can be derived from the modeled straight line of 80% to 20% region of PDD. The same electron energy with different cone sizes was also modeled by the primary-tail function. The stopping power of different electron energies in different depths can also be derived from the parameters , , and .

Results: The main parameters and increase but decreases in the primary-tail function for characterizing the electron beam PDD when the electron energy increased. The relationship of parameter , , and ln(-) with electron energy are = 31.667 - 88, = 0.9975 - 2.8535, and ln(-) = -0.1355 - 6.0986, respectively. Percent depth dose was derived from the percent reading curve by multiplying the stopping power relevant to the depth in water at a certain electron energy. The stopping power of different electron energies can be derived from and with the following equation: stopping power = (-0.042ln( ) + 1.072) , where is the depth in water. The lateral scatter equivalence (LSE) of the clinical electron beam can be described by the parameters , , and in the equation of S = ( - )/( / ). The LSE was compared with the root mean square scatter angular distribution method and shows the agreement of depth dose distributions within ±2%.

Conclusions: The PDD of the electron beam at different energies and cone sizes can be modeled with an empirical model to deal with what is the minimal field size without changing the percent depth dose when approximate LSE is given in centimeters of water.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8519681PMC
http://dx.doi.org/10.1155/2021/3397350DOI Listing

Publication Analysis

Top Keywords

electron beam
24
depth dose
20
primary-tail function
16
pdd electron
16
electron energy
16
electron
14
percent depth
12
electron energies
12
stopping power
12
function
9

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!