Introduction: Monopolar electrocautery is the most common dissection and coagulation tool during transoral robotic surgery (TORS) but causes significant collateral tissue damage as opposed to CO laser. We aimed at combining both modalities in one robotic instrument arm.
Methods: We developed a steerable CO -laser fibre carrier serving as an add-on to the existing Endowrist® monopolar spatula of the Da Vinci Xi. Feasibility and safety were assessed in a preclinical setting.
Results: One radical tonsillectomy with monopolar cautery and three with the instrument prototype were performed in two cadavers by two surgeons. No serious prototype-related intra-operative difficulties were observed. Safe and efficient switching between energy sources proved possible in all simulated intra-operative bleeding events. Prototype use allowed for the identification of the majority of key anatomical structures and was scored favourably on NASA-TLX questionnaires.
Discussion: The reported prototype successfully combines the advantages of CO -laser with the advantages of TORS.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1002/rcs.2342 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!