Purpose: To evaluate the accuracy of single-photon emission computed tomography (SPECT) in diagnosing unilateral condylar hyperplasia (UCH) and to describe the condylar growth of patients with UCH.
Materials And Methods: Using a retrospective study design, patients with UCH who had undergone SPECT and cone-beam computed tomography (CBCT) examinations at the same time were included in the study. We used 3D medical models based on CBCT data as the gold standard. The SPECT results were compared with the model data, and the sensitivity and specificity were calculated. To further describe the condylar growth activity, statistical analysis was performed, and the P value was set at 0.05.
Results: The sample was composed of 75 patients. The sensitivity of SPECT was 55.3%, the specificity was 48.6%, and the area under the receiver operating characteristic curve was 0.53. There was no significant difference in sex between patients with and without active growth.
Conclusion: The sensitivity and specificity of SPECT are poor, and SPECT alone is not suitable for evaluating the active stage of condylar growth. 3D medical modeling has good prospects for application in the diagnosis of condylar hyperplasia.