Purpose: Craniospinal irradiation (CSI) using tomotherapy has advantages over standard 3-dimensional techniques. However, there is a paucity of published data on craniospinal setup reproducibility to guide appropriate planning treatment volume (PTV) margins. We sought to evaluate the setup accuracy of patients undergoing CSI to optimize PTV margins.

Methods And Materials: We measured residual setup deviation between simulation computed tomography (CT) and daily megavoltage CT after couch shifts made by therapists after megavoltage CT-based image registration for 10 patients who completed CSI at our institution. Translational displacement values were recorded at the sella, top of T1, and top of L5 in the anteroposterior (AP) and lateral planes. Systematic and random error were calculated from displacement values. Using score analysis, we calculated minimal PTV margins to encompass 90% of recorded fractions at each level. We evaluated whether patient characteristics predict for increased setup error using standard statistical techniques.

Results: The mean setup deviation in the AP plane across all treatments was 2.49, 3.40, and 3.83 mm at the sella, T1, and L5, respectively. Mean lateral setup error was 2.86, 4.02, and 5.46 mm at the sella, T1, and L5, respectively. Systematic error ranged from 0.75 to 1.01 mm at the sella, 1.09 to 1.37 mm at T1, and 1.30 to 1.50 mm at L5. Random error ranged from 1.35 to 1.41 mm at the sella, 1.48 to 1.73 mm at T1, and 2.26 to 2.37 mm at L5. The minimum margin to cover 90% of the treatments was 6.4, 8.2, and 10.5 mm at the sella, T1, and L5, respectively. There appeared to be a correlation between older age and lateral setup error in the L spine approaching statistical significance (R, 0.629;  = .052).

Conclusions: Setup error increases in the caudal direction of the spine and is greater in the lateral plane compared with the AP plane. We recommend a PTV margin of 5 to 7 mm in the brain and 10 mm in the spine.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8498716PMC
http://dx.doi.org/10.1016/j.adro.2021.100747DOI Listing

Publication Analysis

Top Keywords

setup error
16
setup
9
setup accuracy
8
craniospinal irradiation
8
planning treatment
8
treatment volume
8
ptv margins
8
setup deviation
8
displacement values
8
random error
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!