Background: During cancer operations, the cancer itself is often hard to delineate-buried beneath healthy tissue and lacking discernable differences from the surrounding healthy organ. Long-wave infrared, or thermal, imaging poses a unique solution to this problem, allowing for the real-time label-free visualization of temperature deviations within the depth of tissues. The current study evaluated this technology for intraoperative cancer detection.

Methods: In this diagnostic study, patients with gastrointestinal, hepatobiliary, and renal cancers underwent long-wave infrared imaging of the malignancy during routine operations.

Results: It was found that 74% were clearly identifiable as hypothermic anomalies. The average temperature difference was 2.4°C (range 0.7 to 5.0) relative to the surrounding tissue. Cancers as deep as 3.3 cm from the surgical surface were visualized. Yet, 79% of the images had clinically relevant false positive signals [median 3 per image (range 0 to 10)] establishing an accuracy of 47%. Analysis suggests that the degree of temperature difference was primarily determined by features within the cancer and not peritumoral changes in the surrounding tissue.

Conclusion: These findings provide important information on the unexpected hypothermal properties of intra-abdominal cancers, directions for future use of intraoperative long-wave infrared imaging, and new knowledge about the in vivo thermal energy expenditure of cancers and peritumoral tissue.

Download full-text PDF

Source
http://dx.doi.org/10.1177/15533506211046096DOI Listing

Publication Analysis

Top Keywords

long-wave infrared
16
infrared imaging
12
intraoperative cancer
8
temperature difference
8
cancer
5
long-wave
4
imaging
4
imaging intraoperative
4
cancer detection-what
4
detection-what true
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!