Purpose: Cone beam computed tomography (CBCT) is a standard solution for in-room image guidance for radiation therapy. It is used to evaluate and compensate for anatomopathological changes between the dose delivery plan and the fraction delivery day. CBCT is a fast and versatile solution, but it suffers from drawbacks like low contrast and requires proper calibration to derive density values. Although these limitations are even more prominent with in-room customized CBCT systems, strategies based on deep learning have shown potential in improving image quality. As such, this article presents a method based on a convolutional neural network and a novel two-step supervised training based on the transfer learning paradigm for shading correction in CBCT volumes with narrow field of view (FOV) acquired with an ad hoc in-room system.
Methods: We designed a U-Net convolutional neural network, trained on axial slices of corresponding CT/CBCT couples. To improve the generalization capability of the network, we exploited two-stage learning using two distinct data sets. At first, the network weights were trained using synthetic CBCT scans generated from a public data set, and then only the deepest layers of the network were trained again with real-world clinical data to fine-tune the weights. Synthetic data were generated according to real data acquisition parameters. The network takes a single grayscale volume as input and outputs the same volume with corrected shading and improved HU values.
Results: Evaluation was carried out with a leave-one-out cross-validation, computed on 18 unique CT/CBCT pairs from six different patients from a real-world dataset. Comparing original CBCT to CT and improved CBCT to CT, we obtained an average improvement of 6 dB on peak signal-to-noise ratio (PSNR), +2% on structural similarity index measure (SSIM). The median interquartile range (IQR) Hounsfield unit (HU) difference between CBCT and CT improved from 161.37 (162.54) HU to 49.41 (66.70) HU. Region of interest (ROI)-based HU difference was narrowed by 75% in the spongy bone (femoral head), 89% in the bladder, 85% for fat, and 83% for muscle. The improvement in contrast-to-noise ratio for these ROIs was about 67%.
Conclusions: We demonstrated that shading correction obtaining CT-compatible data from narrow-FOV CBCTs acquired with a customized in-room system is possible. Moreover, the transfer learning approach proved particularly beneficial for such a shading correction approach.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9297981 | PMC |
http://dx.doi.org/10.1002/mp.15282 | DOI Listing |
Plant Cell Physiol
January 2025
Astrobiology Center, National Institutes of Natural Sciences, Mitaka 181-8588, Japan.
Heterogeneous distribution of PSI and PSII in thick grana in shade chloroplasts is argued to hinder spillover of chlorophyll excitations from PSII to PSI. To examine this dogma, we measured fluorescence induction at 77K at 690 nm (PSII) and 760 nm (mostly PSI) in the leaf discs of Spinacia oleracea, Cucumis sativus and shade tolerant Alocasia odora, grown at high and low light, and quantified their spillover capacities. PSI fluorescence (FI) consists of the intrinsic PSI fluorescence (FIα) and fluorescence caused by excitations spilt over from PSII (FIβ).
View Article and Find Full Text PDFClin Oral Investig
January 2025
Center for Dental Medicine, Department of Operative Dentistry and Periodontology, Faculty of Medicine and Medical Center, University of Freiburg, Hugstetter Str. 55, 79106, Freiburg im Breisgau, Germany.
Objectives: The study aimed to assess the percent correct shade identification of four intraoral scanners (IOS) and a spectrophotometer, focusing on how reliably each device selects the correct tooth shade compared to a visual observer's selection. The research question addresses how much clinicians can trust the device-selected shade without visual verification.
Materials And Methods: Sixteen participants with natural, unrestored teeth were included.
Brain Struct Funct
December 2024
Institute of Movement and Neurosciences, Department of Exercise Neuroscience, German Sport University Cologne, Am Sportpark Müngersdorf 6, Cologne, D-50933, Germany.
Med Phys
December 2024
Research Center for Advanced Detection Materials and Medical Imaging Devices, Shenzhen Institute of Advanced Technology, Chinese Academy of Sciences, Shenzhen, Guangdong, China.
Background: Recently, the popularity of dual-layer flat-panel detector (DL-FPD) based dual-energy cone-beam CT (CBCT) imaging has been increasing. However, the image quality of dual-energy CBCT remains constrained by the Compton scattered x-ray photons.
Purpose: The objective of this study is to develop a novel scatter correction method, named e-Grid, for DL-FPD based CBCT imaging.
Materials (Basel)
December 2024
Department of Restorative Dentistry, Faculty of Dental Medicine, Hokkaido University, Kita 13 Nishi 7, Kita-ku, Sapporo 060-8586, Japan.
Background: This study aimed to evaluate the color-matching and light transmission properties of a newly developed aesthetic flowable resin composite, OCFB-001.
Methods: Rubber molds containing cylindrical cavities were filled with Estelite Sigma Quick, and 40 resin composite (CR) molds with simulated Class I cavities were prepared in shades A1, A2, A3, and A4, resulting in a total of 160 samples. Following bonding procedures, four different flowable resin composites ( = 10) were introduced into the cavities.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!