To demonstrate the large potential of proton minibeam radiotherapy (pMBRT) as a new method to treat tumor diseases, a preclinical proton minibeam radiation facility was designed. It is based on a tandem Van-de-Graaff accelerator providing a 16 MeV proton beam and a 3 GHz linac post-accelerator (designs: AVO-ADAM S.A, Geneva, Switzerland and ENEA, Frascati, Italy). To enhance the transmission of the tandem beam through the post-accelerator by a factor of 3, two drift tube buncher units were designed and constructed: A brazed 5-gap structure (adapted SCDTL tank of the TOP-IMPLART project (ENEA)) and a non-brazed low budget 4-gap structure. Both are made of copper. The performance of the two differently manufactured units was evaluated using a 16 MeV tandem accelerator beam and a Q3D magnetic spectrograph. Both buncher units achieve the required summed voltage amplitude of 42 kV and amplitude stability at a power feed of less than 800 W.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8504737PMC
http://journals.plos.org/plosone/article?id=10.1371/journal.pone.0258477PLOS

Publication Analysis

Top Keywords

buncher units
12
proton minibeam
12
preclinical proton
8
concept performance
4
performance evaluation
4
evaluation ghz
4
ghz buncher
4
units
4
units optimizing
4
optimizing dose
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?