Objectives: We assessed the impact of conventional delivery system (DS) insertion technique on "Hat-marker" orientation/commissural alignment in patients who underwent transcatheter aortic valve replacement (TAVR) in the Evolut Low Risk Trial CT substudy versus a modified technique.

Background: Unlike surgical aortic valve replacement, where alignment of the surgical valve commissures with native commissures can be achieved virtually 100% of the time, commissural alignment during TAVR is not achieved consistently. This may subsequently impact the feasibility of both coronary access and reintervention after TAVR.

Methods: "Hat-marker" orientations during deployment were characterized as outer curve (OC), center front (CF), inner curve, and center back. Severe commissure-to-CA overlap was 0-20°. "Hat-marker" orientations and CA overlap were compared to 240 patients from a single center using the modified 3-o'clock flush port DS technique.

Results: In the CT substudy in which conventional DS insertion was performed (flush port at 12 o'clock); 154/249 had both analyzable CT and procedural fluoroscopy to validate "Hat-marker" to C-tab/commissural orientation. On post-TAVR CT, Evolut valve commissural orientation and coronary artery (CA) ostia were identified. Compared to conventional DS technique in the CT substudy, the modified technique had higher rates of "Hat-marker" at OC/CF orientation, improved commissural alignment and reduced severe CA overlap; (left main, 14.2 vs. 27.9%; right coronary artery, 11.7 vs. 27.3% both, 5.0 vs. 13.6%; 1 or both CA, 20.8 vs. 41.6%, all p < 0.01).

Conclusions: The modified technique improved initial "Hat-marker" orientation during Evolut deployment and resulted in better commissural alignment and reduced CA overlap.

Download full-text PDF

Source
http://dx.doi.org/10.1002/ccd.29973DOI Listing

Publication Analysis

Top Keywords

commissural alignment
12
versus modified
8
delivery system
8
aortic valve
8
valve replacement
8
"hat-marker" orientations
8
curve center
8
flush port
8
coronary artery
8
alignment
5

Similar Publications

Introduction And Objectives: This study evaluated commissural malalignment on echocardiography as a predictor of coronary anomalies.

Methods: All newborns diagnosed with transposition of great arteries in the pediatric cardiac intensive care unit between 1 August 2020 and 1 February 2022 were included in this study. The ratio of distances (C-ratio) from the anterior commissure to the right-sided commissure of the pulmonary valve and the distance from the anterior commissure to the left sided commissure of the pulmonary valve were calculated.

View Article and Find Full Text PDF

Initial Multicenter Experience With a Novel Self-Expanding TAVR System in Patients With Aortic Valve Stenosis.

JACC Cardiovasc Interv

January 2025

Clinic for General and Interventional Cardiology/Angiology, Herz- und Diabeteszentrum NRW, Ruhr-Universität Bochum, Bad Oeynhausen, Germany. Electronic address:

Background: As transcatheter aortic valve replacement is performed increasingly in younger, low-risk patients, the need for commissural alignment and coronary access has increased. Design elements of the JenaValve Trilogy (JVT) transcatheter heart valve (THV) ensure both.

Objectives: This study sought to evaluate the outcome of patients with aortic stenosis (AS) treated with this novel transfemoral, self-expanding THV.

View Article and Find Full Text PDF

Brain color-coded diffusion imaging: Utility of ACPC reorientation of gradients in healthy subjects and patients.

Comput Methods Programs Biomed

December 2024

Université Clermont Auvergne, CNRS, CHU Clermont-Ferrand, Clermont Auvergne INP, Institut Pascal, F-63000 Clermont-Ferrand, France; Université Clermont Auvergne, Clermont Auvergne INP, CNRS, Institut Pascal, F-63000 Clermont-Ferrand, France. Electronic address:

Background And Objective: The common structural interpretation of diffusion color-encoded (DCE) maps assumes that the brain is aligned with the gradients of the MRI machine. This is seldom achieved in the field, leading to incorrect red (R), green (G) and blue (B) DCE values for the expected orientation of fiber bundles. We studied the virtual reorientation of gradients according to the anterior commissure - posterior commissure (ACPC) system on the RGB derivatives.

View Article and Find Full Text PDF

We present one of the inaugural transcatheter aortic valve implantation procedures using the latest Medtronic platform, Evolut FX+. Successful coronary angiography was achieved within 28 seconds and 1 minute 49 seconds for the left and right coronary arteries, respectively. Postoperative cardiac computed tomography scan demonstrated optimal commissural and diamond-coronary alignment.

View Article and Find Full Text PDF
Article Synopsis
  • The study aimed to assess the hemodynamic effects of commissural alignment (CA) in patients receiving self-expandable transcatheter aortic valves (TAVR) using two different devices: Evolut Pro (EP) and Acurate neo2 (AN2).
  • Analyzing 557 patients, the results showed that CA was linked to a significant reduction in the rate of commissural misalignment (CMA) compared to those without CA, leading to lower transvalvular gradients and less aortic regurgitation after one year.
  • Although patients with moderate/severe CMA experienced worsened aortic gradients and regurgitation rates, overall structural deterioration criteria were not met at