A computational multilayer model to simulate hollow needle insertion into biological porcine liver tissue.

Acta Biomater

School of Engineering, Faculty of Science and Engineering, University of Limerick, Limerick, Ireland; BioScience and BioEngineering Research (BioSciBer), Bernal Institute, University of Limerick, Ireland; Health Research Institute, University of Limerick, Limerick, Ireland. Electronic address:

Published: December 2021

Modelling of needle insertion in soft tissue has developed significant interest in recent years due to its application in robot-assisted minimally invasive surgeries such as biopsies and brachytherapy. However, this type of surgery requires real-time feedback and processing which complex computational models may not be able to provide. In contrast to the existing mechanics-based kinetic models, a simple multilayer tissue model using a Coupled Eulerian Lagrangian based Finite Element method has been developed using the dynamic principle. The model simulates the needle motion for flexible hollow bevel-angled needle (15° and 30°, 22 Gauge) insertion into porcine liver tissue, which includes material parameters obtained from unconfined compression testing of porcine liver tissue. To validate simulation results, needle insertion force and cutting force within porcine liver tissue were compared with corresponding experimental results obtained from a custom-built needle insertion system. For the 15° and 30° bevel-angle needles, the percentage error for cutting force (mean) of each needle compared to computational model, were 18.7% and 11.9% respectively. Varying the needle bevel angle from 30° to 15° results in an increase of the cutting force, but insertion force does not vary among the tested bevel angles. The validation of this computationally efficient multilayer Finite Element model can help engineers to better understand the biomechanical behaviour of medical needle inside soft biological tissue. Ultimately, this multilayer approach can help advance state-of-art clinical applications such as robot-assisted surgery that requires real-time feedback and processing. STATEMENT OF SIGNIFICANCE: The significance of the work is in confirming the effectiveness of multilayer material based finite element (FE) method to model biopsy needle insertion into soft biological porcine liver tissue. A multilayer Coupled Eulerian Lagrangian (CEL) based FE modelling technique allowed testing of heterogeneous, non-linear viscoelastic porcine liver tissue in a system, so direct comparison of needle tissue interaction forces on the intrinsic material (tissue) behaviour could be made. To the best of the authors' knowledge, the present research investigates for the first time modelling of a three dimensional (3D) hollow needle insertion using a multilayer stiffness model of biological tissue using FE based CEL method and presents a comparison of simulation results with experimental data.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.actbio.2021.09.057DOI Listing

Publication Analysis

Top Keywords

needle insertion
24
porcine liver
24
liver tissue
24
needle
12
tissue
12
finite element
12
cutting force
12
hollow needle
8
insertion
8
biological porcine
8

Similar Publications

Optogenetics has transformed the study of neural circuit function, but limitations in its application to species with large brains, such as non-human primates (NHPs), remain. A major challenge in NHP optogenetics is delivering light to sufficiently large volumes of deep neural tissue with high spatiotemporal precision, without simultaneously affecting superficial tissue. To overcome these limitations, we recently developed and tested in NHP cortex, the Utah Optrode Array (UOA).

View Article and Find Full Text PDF

What Is the Optimal Geometry of Dissolving Microneedle Arrays? A Literature Review.

Pharmaceutics

January 2025

Department of Pharmaceutical Technology and Biopharmacy, University of Groningen, 9713 AV Groningen, The Netherlands.

The application of dissolving microneedle arrays (DMNAs) is an emerging trend in drug and vaccine delivery as an alternative for hypodermic needles or other less convenient drug administration methods. The major benefits include, amongst others, that no trained healthcare personnel is required and that the recipient experiences hardly any pain during administration. However, for a successful drug or vaccine delivery from the DMNA, the microneedles should be inserted intact into the skin.

View Article and Find Full Text PDF

Reduce electrical overload via threaded Chinese acupuncture in nerve electrical therapy.

Bioact Mater

April 2025

Department of Orthopaedics, Shanghai Key Laboratory for Prevention and Treatment of Bone and Joint Diseases, Shanghai Institute of Traumatology and Orthopaedics, Ruijin Hospital, Shanghai Jiao Tong University School of Medicine, 197 Ruijin 2nd Road, Shanghai, 200025, PR China.

Bioelectrical stimulation is a powerful technique used to promote tissue regeneration, but it can be hindered by an "electrical overload" phenomenon in the core region of stimulation. We develop a threaded microneedle electrode system that protects against "electrical overload" by delivering medicinal hydrogel microspheres into the core regions. The threaded needle body is coated with polydopamine and chitosan to enhance the adhesion of microspheres, which are loaded into the threaded grooves, allowing for their stereoscopic release in the core regions.

View Article and Find Full Text PDF

Objective: To investigate the clinical efficacy of percutaneous prying combined with modified rotary reduction with needle in the treatment of supracondylar fracture of humerus in Gartland type Ⅳ children, and to evaluate the postoperative elbow joint function, the incidence of elbow varus deformity and the application prospect of this technique.

Methods: A total of 98 children diagnosed with Gartland type Ⅳ supracondylar humeral fractures between June 2020 and January 2023 were included in this study, comprising of 57 males and 41 females. The age ranged from 2 to 14 years old with an average of (6.

View Article and Find Full Text PDF

Objective: To discuss the elbow skin fold extension line in Kirschner wire internal fixation of extended supracondylar humeral fractures in children.

Methods: The clinical data of 58 children with extended supracondylar fractures of the humerus who met the selection criteria between August 2021 and July 2024 were retrospectively analyzed. In 28 cases, needle placement of medial epicondyle of humerus was performed with the assistance of the elbow skin fold extension line (study group), and 30 cases were assisted by routine touch of the medial epicondyle of the humerus (control group).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!