COMPARISON OF SMARTPHONE-BASED AND AUTOMATED REFRACTION WITH SUBJECTIVE REFRACTION FOR SCREENING OF REFRACTIVE ERRORS.

Ophthalmic Epidemiol

University of Malaya Eye Research Centre, Department of Ophthalmology, Faculty of Medicine, University of Malaya, Kuala Lumpur, Malaysia.

Published: October 2022

Purpose: To compare Netra smartphone-based and automated refraction with subjective refraction for screening of refractive errors.

Methods: Cross-sectional study at the University of Malaya Medical Centre, Kuala Lumpur. Subjects underwent subjective refraction, then automated refraction, and finally Netra smartphone-based refraction. All results were converted to power vectors (M, J0 and J45) and were analysed using repeated-measures ANOVA and Bland-Altman plots. Sensitivity and specificity were determined. The best cut-off points were determined from ROC curve analysis. < .05 was considered statistically significant.

Results: Data from the right eyes of 204 subjects were analysed. Mean age was 36.6 ± 15.7 years (range 16-78 years). Spherical equivalent [mean (95% CI)] from Netra and automated refraction were similar, and both more myopic than subjective refraction; -2.87 (-3.23 to -2.51), -2.85 (-3.21 to -2.49) and -2.46 (-2.83 to -2.10) respectively ( < .001). Differences in J0 and J45 between Netra and subjective refraction were not statistically significant (0.10 vs 0.11 and 0.01 vs -0.02 respectively, both > .05), but those between automated and subjective refraction were (0.06 vs 0.11 and 0.07 vs -0.02, = .004 and < .001 respectively). Bland Altman plots showed the 95% limits of agreement with Netra refraction were wider than with automated refraction (-2.21D to 1.42D vs. -1.90D to 1.16D respectively).

Conclusion: Netra smartphone-based refraction gives similar readings to automated refraction, and both show myopic overestimation when compared to subjective refraction. However, due to non-insignificant practical usage issues, its use as a screening tool for refractive errors is limited.

Download full-text PDF

Source
http://dx.doi.org/10.1080/09286586.2021.1986550DOI Listing

Publication Analysis

Top Keywords

automated refraction
12
subjective refraction
12
smartphone-based automated
8
refraction subjective
8
refraction screening
8
screening refractive
8
netra smartphone-based
8
refraction
7
comparison smartphone-based
4
refractive errors
4

Similar Publications

Optical diffraction tomography using a self-reference module.

Biomed Opt Express

January 2025

Department of Electronic Engineering, Maynooth University, Maynooth, Co. Kildare, Ireland.

Optical diffraction tomography enables label-free, 3D refractive index (RI) imaging of biological samples. We present a novel, cost-effective approach to ODT that employs a modular design incorporating a self-reference holographic capture module. This two-part system consists of an illumination module and a capture module that can be seamlessly integrated with any life-science microscope using an automated alignment protocol.

View Article and Find Full Text PDF

Purpose: The diagnosis of fungal keratitis using potassium hydroxide (KOH) smears of corneal scrapings enables initiation of the correct antimicrobial therapy at the point-of-care but requires time-consuming manual examination and expertise. This study evaluates the efficacy of a deep learning framework, dual stream multiple instance learning (DSMIL), in automating the analysis of whole slide imaging (WSI) of KOH smears for rapid and accurate detection of fungal infections.

Design: Retrospective observational study.

View Article and Find Full Text PDF

Significance: Optimal meibography utilization and interpretation are hindered due to poor lid presentation, blurry images, or image artifacts and the challenges of applying clinical grading scales. These results, using the largest image dataset analyzed to date, demonstrate development of algorithms that provide standardized, real-time inference that addresses all of these limitations.

Purpose: This study aimed to develop and validate an algorithmic pipeline to automate and standardize meibomian gland absence assessment and interpretation.

View Article and Find Full Text PDF

As one of the typical applications of metamaterials, the invisibility cloak has raised vast research interests. After many years' research efforts, the invisibility cloak has extended its applicability from optics and acoustics to electrostatics and thermal diffusion. One scientific challenge that has significantly restricted the practical application of the invisibility cloak is the strong background dependence, that is, all passive cloaking devices realized thus far are unable to resist variation in the background refractive index.

View Article and Find Full Text PDF

Purpose: To quantify outer retina structural changes and define novel biomarkers of inherited retinal degeneration associated with biallelic mutations in RPE65 (RPE65-IRD) in patients before and after subretinal gene augmentation therapy with voretigene neparvovec (Luxturna).

Methods: Application of advanced deep learning for automated retinal layer segmentation, specifically tailored for RPE65-IRD. Quantification of five novel biomarkers for the ellipsoid zone (EZ): thickness, granularity, reflectivity, and intensity.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!