Background: To study and compare the clinical optical image quality following implantation with different premium IOLs by analysing the point spread function (PSF) Strehl ratio using a pyramidal wavefront sensor (PWS)-based aberrometer.
Methods: This study included 194 eyes implanted with: (a) 19 AcrySof SA60AT (control group); (b) 19 Miniwell; (c) 24 LENTIS Mplus LS-313 MF30; d) 33 LENTIS Mplus LS-313 MF15; (e) 17 AkkoLens Lumina; (f) 31 AT LISA Tri 839MP; (g) 20 Precizon Presbyopic; (h) 20 AcrySof IQ PanOptix; (i) 11 Tecnis Eyhance. Main outcome measures were PSF Strehl ratio, PSF Strehl ratio excluding second-order aberrations (PSFw2), total root mean square (RMS), low-order aberration (LOA) and high-order aberration (HOA) RMS measured by PWS aberrometer.
Results: AT LISA Tri had the highest PSFw2 Strehl ratio at both 3.0- and 4.0-mm pupil size (0.52 ± 0.14 and 0.31 ± 0.10; P < 0.05), followed by SA60AT (0.41 ± 0.11 and 0.28 ± 0.07) and PanOptix (0.4 ± 0.07 and 0.26 ± 0.04). AT LISA Tri was found to provide a significantly better retinal image quality than PanOptix at both 3.0 mm (P < 0.0001) and 4.0 mm (P = 0.004). Mplus MF15 was found to be significantly better than Mplus MF30 at both 3.0 mm (P < 0.0001) and 4.0 mm (P = 0.002). Total RMS, LOA RMS, HOA RMS, PSF Strehl ratio and PSFw2 varied significantly between the studied groups (P < 0.001).
Conclusions: Far distance clinical image quality parameters measured by PWS aberrometer differed significantly according to the technology of the implanted lens. AT LISA Tri, SA60AT and PanOptix showed the highest values of far distance retinal image quality, while the lowest PSFw2 Strehl ratios were displayed by Miniwell, Mplus MF30 and Precizon Presbyopic.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8496005 | PMC |
http://dx.doi.org/10.1186/s40662-021-00258-y | DOI Listing |
Eye Contact Lens
November 2024
Tianjin Key Laboratory of Retinal Functions and Diseases, Tianjin Branch of National Clinical Research Center for Ocular Disease, Eye Institute and School of Optometry, Tianjin Medical University Eye Hospital, Tianjin, China.
Objectives: To evaluate the impact of scleral contact lens (SL) wear on the visual quality and the ocular surface wettability in myopic patients with regular corneas.
Methods: This prospective, randomized, controlled study enrolled a total of 80 myopes with regular corneas. Subjects were randomly allocated to wear SL or rigid corneal lens (RCL) for 3 months.
Zhonghua Yan Ke Za Zhi
January 2025
Eye Institute of Shandong First Medical University, Qingdao Eye Hospital of Shandong First Medical University, State Key Laboratory Cultivation Base, Shandong Key Laboratory of Eye Diseases, School of Ophthalmology, Shandong First Medical University, Qingdao266071, China.
To evaluate the safety and efficacy of cataract extraction combined with multifocal intraocular lens (IOL) optic implantation in Berger space in school-age children with cataracts. It was a prospective study. The clinical data of school-age children with cataract who underwent cataract extraction combined with multifocal IOL implantation at Qingdao Eye Hospital of Shandong First Medical University from January 2019 to June 2023 were collected.
View Article and Find Full Text PDFPurpose: To compare the aberrometric profile of the Artisan Aphakia lens (Opthec BV) and transscleral plug FIL-SSF lens (Soleko) and to assess the impact of tilt and decentration on their optical performance.
Methods: This retrospective observational study was conducted at Azienda Ospedaliera Universitaria Careggi, Florence, Italy, with a consecutive cohort of aphakic eyes undergoing secondary lens implantation with an Artisan or FIL-SSF lens. Wavefront analysis was performed using a pyramidal wavefront-based aberrometer (Osiris-T; CSO).
PLoS One
December 2024
UFR, Department of Ophthalmology, University Hospital Miguel Servet, Zaragoza, Spain.
Purpose: Accurate alignment of Toric Intraocular Lens (T-IOLs) in cataract surgery is crucial for good visual outcomes. The purpose of this study was to evaluate the influence of rotation, axial shift and their combined effects on the refractive error and image quality of a wide range of T-IOL powers (from +1.50 D to +6.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!