Evaluation of the Accuracy of Standardized Uptake Values of F-fluorodeoxyglucose in Lung Lesions Based on Phantom Studies.

Sovrem Tekhnologii Med

Researcher, Laboratory of Radiation Hygiene of Medical Facilities; Saint Petersburg Research Institute of Radiation Hygiene named after Professor P.V. Ramzaev, 8 Mira St., Saint Petersburg, 197101, Russia; Researcher; A.M. Granov Russian Research Center for Radiology and Surgical Technologies, Ministry of Health of the Russian Federation, 70 Leningradskaya St., Saint Petersburg, Pesochniy pos., 197758, Russia; Associate Professor, Department of Nuclear Medicine and Radiation Technologies, Almazov National Medical Research Centre, Ministry of Health of the Russian Federation, 2 Akkuratova St., Saint Petersburg, 197341, Russia.

Published: March 2022

Unlabelled: was to estimate the accuracy of standardized uptake values of F-fluorodeoxyglucose (F-FDG) in lung lesions during positron emission tomography combined with computed tomography (PET/CT) imaging, based on phantom studies performed for different PET/CT scanners.

Materials And Methods: The analysis of the PET/CT with F-FDG data was performed for 86 patients newly diagnosed with the lung lesions: malignant tumors (n=37), benign tumors and inflammatory diseases (n=49). The criteria for inclusion in the study were developed considering the recommendations of the Fleischner Society (2017). The characteristics of the lesions on CT met the following requirements: a round shape or close to it; total size of 8 to 30 mm; solid or subsolid structure (with the exception of lesion with ground-glass opacity); a solid part size of ≥8 mm. All the patients had no signs of pleurisy, lymphadenopathy, or cancer history. PET/CT imaging with F-FDG was performed with three scanners: Discovery 690 (General Electric, USA), Biograph mCT 128 (Siemens, Germany), and Biograph mCT 40 (Siemens); the preparation of patients prior to the scan was standardized. To determine the reference accumulation of a radiopharmaceutical in the pathological lesion, four scans of a specialized NEMA IEC PET Body Phantom Set (USA) were performed for each scanner. For each unit, the recovery coefficients (RCs) of radioactivity, maximum and recovered (corrected) standardized uptake values (SUVs) were determined. Statistical relationship between the size of lesions, SUV and SUV was evaluated. Data processing was performed using MedCalc v. 19.2.0 software.

Results: During the phantom study, the underestimation of the radioactivity was determined in the spheres with the diameters of 10 and 13 mm, overestimation was observed in the sphere with the diameter of 28 mm. Both underestimation and overestimation of radioactivity were determined for the spheres with a diameter of 17 and 22 mm.SUV differed from the reference values for 85 patients (98.8%). The underestimation of these values was found for 63 patients (73.2%) due to the partial volume effect. The greatest underestimation was observed for the patients with 8 mm diameter lesions. Depending on the scanner, the underestimation of the SUV in these patients reached up to 54-73%. For 9 patients (25%) with malignant tumors of 9-12 mm, the utility of RC made it possible to avoid false negative results. For the lesions with a diameter of 30 mm, an overestimation of SUV up to 22% was determined due to the negative influence of the reconstruction algorithms.

Conclusion: The use of RC eliminates the influence of the partial volume effect and reconstruction methods on the accuracy of estimating the SUV in lung lesions, which ensures reproducibility, increase in the information content of the method, as well as the comparability of the results of PET/CT with F-FDG obtained on the different models of PET/CT units with different technological characteristics.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8482825PMC
http://dx.doi.org/10.17691/stm2021.13.3.02DOI Listing

Publication Analysis

Top Keywords

lung lesions
16
standardized uptake
12
uptake values
12
accuracy standardized
8
values f-fluorodeoxyglucose
8
lesions
8
based phantom
8
phantom studies
8
pet/ct imaging
8
pet/ct f-fdg
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!