Objectives: Implant failure leading to revision total joint arthroplasty can occur through a variety of different mechanisms which are typically associated with a soft tissue response adjacent to the implant that provide insight into the underlying etiology of implant failure. The objective of this study was to elucidate mechanisms of implant failure as they relate to histological classification and findings of adjacent periprosthetic tissue.

Methods: Histological analysis of soft tissue adjacent to the implant was performed in 99 patients with an average age of 64 years old and grouped into four categories based on the study conducted by Morawietz et al.:Type I (N = 47)Wear particle induced typeType II (N = 7)Infectious typeType III (N = 19)Combined type I and IIType IV (N = 26)Indeterminant typeModes of failure were categorized into five groupings based on the study conducted by Callies et al.: Instability (N = 35), Aseptic Loosening (N = 24), Hardware and/or Mechanical Failure (N = 15), Septic (N = 13), and Other failures (N = 12). We calculated odds ratios and conducted regression analysis to assess the relationship between modes of failure and histological findings as well as modes of failure and comorbidities.

Results: Hardware/mechanical failure was independently correlated with histological findings of anucleate protein debris, histiocytes, , and synovitis. Furthermore, hardware/mechanical failure was independently correlated with osteosarcoma as a co-morbidity. Septic failure was associated with histological findings of Enterococcus, granulation tissue, and tissue necrosis as well as comorbidities of Crohn's disease, deep venous thrombosis, lung disease, and rheumatoid arthritis. Infection was 5.8 times more likely to be associated with Type II histology. Aseptic loosening was associated with histologic findings of synovitis.

Conclusion: Our findings support the existing literature on periprosthetic tissue analysis in revision total joint arthroplasty which may improve surgeon understanding of the patholophysiological mechanisms that contribute to implant failure and revision surgery.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8474343PMC
http://dx.doi.org/10.1177/20503121211047099DOI Listing

Publication Analysis

Top Keywords

implant failure
16
soft tissue
12
failure
12
modes failure
12
revision total
12
total joint
12
joint arthroplasty
12
histological findings
12
failure revision
8
adjacent implant
8

Similar Publications

Background: The most effective way to treat patients following a first ICD therapy is unclear. We hypothesised that following first ICD therapy, combining different treatment strategies would be associated with a reduction in the risk of subsequent therapy compared to single strategies alone.

Methods: Data was collected from consecutive patients undergoing ICD implantation at King's College Hospital between January 2009 and December 2019.

View Article and Find Full Text PDF

Background: Accurate prediction for survival in individualized patients with cardiac resynchronization therapy with a defibrillator (CRT-D) is difficult.

Methods: We analyzed the New Japan cardiac device treatment registry (JCDTR) database to develop a survival prediction model for CRT-D recipients.

Results: Four hundred and eighty-two CRT-D recipients, at the implantation year 2018-2021, with a QRS width ≥120 ms and left ventricular ejection fraction (LVEF) ≤35% at baseline, were analyzed.

View Article and Find Full Text PDF

Objectives: Transcatheter edge-to-edge repair (TEER) is an alternative for patients with severe degenerative mitral regurgitation (MR). The objective of this study was to compare the outcomes of surgery and TEER in older patients with degenerative MR patients using real life data.

Methods: Consecutives older patients (≥ 65 years-old), with severe symptomatic, degenerative MR requiring surgery or TEER between 2013 and 2023 were included.

View Article and Find Full Text PDF

Hemodynamic Valve Deterioration After Transcatheter Aortic Valve Replacement: Incidence, Predictors, and Clinical Outcomes.

JACC Cardiovasc Interv

January 2025

Department of Cardiology, Inselspital, Bern University Hospital, University of Bern, Bern, Switzerland. Electronic address:

Background: Reports on the durability of transcatheter aortic valve replacement (TAVR) prostheses are scarce and confounded by varying definitions and competing risks of death.

Objectives: The authors sought to determine the incidence, predictors, and clinical outcomes of hemodynamic valve deterioration (HVD) according to the Valve Academic Research Consortium 3 definition after TAVR.

Methods: We analyzed consecutive patients undergoing TAVR in the prospective Bern TAVI (Transcatheter Aortic Valve Implantation) registry between August 2007 and June 2022 for the incidence and predictors of HVD and performed case control-matching to compare outcomes according to HVD.

View Article and Find Full Text PDF

Background: First-generation bioresorbable scaffolds (BRS) increased risks of stent thrombosis and adverse events. The Bioheart scaffold is a new poly-L-lactic acid-based BRS.

Objectives: This study sought to evaluate the efficacy and safety of the BRS in patients with coronary artery disease.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!