Background: Minimal invasive laparoscopic resection of liver tumors is less traumatic compared with open surgical resection and may be a better option for many patients. However, localization of intrahepatic tumors remains a challenge. Availability of hybrid operating rooms, equipped for high performance radiologic imaging, allows for new methods of surgical navigation.

Methods: Twelve patients planned for laparoscopic resection of liver tumors were included. Before resection started, tumors were marked with radiopaque fiducials. Four fiducials were positioned with ultrasound within 1 cm of the tumor. Tumor and fiducials were localized with contrast enhanced cone beam computed tomography. Fluoroscopy with an overlay of cone beam computed tomography markings was projected side-by-side on the same screen as the laparoscopic view to visualize tumor location. The fiducials were eventually removed. Laparoscopic ultrasound, the standard method of localizing a tumor, was also used. The benefits of the 2 visualization methods were estimated by the operator. Procedure times, radiation doses and resection margins were recorded.

Results: Fluoroscopy with radiopaque fiducials provided valuable information, complementing the laparoscopic ultrasound, particularly during the early phase of resection. In the later phase, mobilization of the tumor-containing liver segment caused significant displacement of the fluoroscopic overlay. The technique evolved during course of the study, with decreasing procedure times and radiation doses. Radical resection was achieved for all patients.

Conclusions: Radiopaque fiducials and fluoroscopy can complement laparoscopic ultrasound for guiding resection of liver tumors. Combining radiologic and optical imaging in a hybrid operating suit may facilitate development of augmented reality techniques for surgical navigation.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8812418PMC
http://dx.doi.org/10.1097/SLE.0000000000000991DOI Listing

Publication Analysis

Top Keywords

radiopaque fiducials
16
resection liver
16
liver tumors
16
laparoscopic resection
12
laparoscopic ultrasound
12
resection
9
hybrid operating
8
cone beam
8
beam computed
8
computed tomography
8

Similar Publications

Article Synopsis
  • * Methods: Patients with low-risk and favorable intermediate-risk prostate cancer received a hydrogel spacer and fiducial markers before receiving targeted radiation therapy, with various questionnaires used to assess quality of life before and after treatment.
  • * Results: In a small group of 20 patients, treatment showed minor positive and negative impacts on urinary, bowel, sexual, and hormonal functions, with low rates of side effects, while PSA levels decreased on average, indicating a potential treatment benefit.
View Article and Find Full Text PDF

Introduction: Precise patient positioning with image guidance (IGRT) is essential for safe prostate radiotherapy. We present the first report of utilizing a CT-visible hydrogel spacer, used to decrease rectal radiation dose, as a surrogate fiducial marker to aid in daily IGRT with cone-beam CT (CBCT) in stereotactic radiotherapy (SABR) for prostate cancer.

Materials And Methods: Prior to CT simulation, patients underwent placement of three intraprostatic gold fiducial markers and radiopaque hydrogel spacer per standard practice.

View Article and Find Full Text PDF

Background: Dose-escalated radiotherapy is known to improve progression free survival in patients with localized prostate cancer, and recent advances have led to the standardization of ultrahypofractionated stereotactic ablative radiotherapy (SABR) delivered in just 5-fractions. Based on the known effectiveness of the accepted though invasive 2-fraction treatment method of high-dose-rate brachytherapy and given the ubiquity of prostate cancer, a further reduction in the number of treatments of external-beam SABR is possible. This study aims to evaluate the safety, efficacy, and non-inferiority of generalizable 2-fraction SABR compared to the current 5-fraction regimen.

View Article and Find Full Text PDF

An open-source MRI compatible frame for multimodal presurgical mapping in macaque and capuchin monkeys.

J Neurosci Methods

July 2024

Department of Bioengineering, University of Pittsburgh, Pittsburgh, PA, USA; Department of Neurobiology, University of Pittsburgh, Pittsburgh, PA, USA; Aligning Science Across Parkinson's (ASAP) Collaborative Research Network, Chevy Chase, MD, USA. Electronic address:

Background: High-precision neurosurgical targeting in nonhuman primates (NHPs) often requires presurgical anatomical mapping with noninvasive neuroimaging techniques (MRI, CT, PET), allowing for translation of individual anatomical coordinates to surgical stereotaxic apparatus. Given the varied tissue contrasts that these imaging techniques produce, precise alignment of imaging-based coordinates to surgical apparatus can be cumbersome. MRI-compatible stereotaxis with radiopaque fiducial markers offer a straight-forward and reliable solution, but existing commercial options do not fit in conformal head coils that maximize imaging quality.

View Article and Find Full Text PDF

Single scanning of CBCT and intraoral scanning for guided implantation in terminal dentitions with multi-unit metal restorations: technical note.

J Stomatol Oral Maxillofac Surg

December 2024

Department of Prosthodontics, State Key Laboratory of Oral & Maxillofacial Reconstruction and Regeneration, Key Laboratory of Oral Biomedicine Ministry of Education, Hubei Key Laboratory of Stomatology, School & Hospital of Stomatology, Wuhan University, Wuhan, PR China. Electronic address:

When anatomical landmarks are missing or obstructed by metal artefacts, it is challenging to accurately merge cone beam computed tomography (CBCT) and intraoral scanning (IOS) information, and the accuracy of the implant surgical guides would be compromised. This article describes a novel technical note using oral wound dressings and flowable resin as additional new radiopaque fiducial landmarks to design surgical guides for full-arch immediate implant placement. This technical note provided an accurate, convenient, and cost-effective option for the clinician.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!