Investigation of attachment location of adhesive skin electrodes for intraoperative neuromonitoring in thyroid surgery: Preclinical and clinical studies.

Surgery

Department of Otorhinolaryngology, Head and Neck Surgery, Pusan National University School of Medicine, Pusan National University and Biomedical Research Institute, Pusan National University Hospital, Busan, Republic of Korea. Electronic address:

Published: February 2022

Background: Recently, adhesive skin electrodes have been reported to be useful for recording electromyographic signals from intrinsic laryngeal muscles for intraoperative neuromonitoring in thyroid surgery and have shown good results compared to existing recording methods. In this study, we investigated the optimal attachment location of adhesive skin electrodes for intraoperative neuromonitoring in both porcine models and human cases.

Methods: Attachment locations were divided vertically into upper, middle, and lower locations and horizontally into medial and lateral locations to determine the optimal location of placing adhesive skin electrodes preclinically in four porcine models. This study included a total of 78 patients who underwent thyroidectomy under intraoperative neuromonitoring with adhesive skin electrodes. Sixteen patients were monitored using both adhesive skin electrodes and an electromyographic endotracheal tube. Two pairs of skin electrodes were attached to the level of the thyroid cartilage lamina. Evoked electromyographic data, including data on mean amplitude and latency, obtained by stimulating the recurrent laryngeal nerve and vagus nerve, were collected.

Results: Lateral attachment of adhesive skin electrodes showed significantly higher evoked amplitudes than medial attachment in both animal models and human patients. In cases where skin electrodes and an electromyographic endotracheal tube were used together, the electromyographic endotracheal tube showed a significantly higher amplitude than skin electrodes, and laterally attached skin electrodes showed a significantly higher amplitude than medially attached skin electrodes.

Conclusion: Intraoperative neuromonitoring using adhesive skin electrodes was feasible in both animal models and human patients. We suggest that it would be better to attach adhesive skin electrodes to the lateral side of the thyroid cartilage lamina. Lateral attachment closer to the cricoarytenoid joint may be better for measuring muscle movement around the cricoarytenoid joint.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.surg.2021.07.037DOI Listing

Publication Analysis

Top Keywords

skin electrodes
52
adhesive skin
36
intraoperative neuromonitoring
20
skin
14
electrodes
13
models human
12
electromyographic endotracheal
12
endotracheal tube
12
adhesive
9
attachment location
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!