At present, therapeutic drug monitoring is the standard in pharmacotherapy using medications with a narrow therapeutic index or showing serious adverse effects, such as in the case of ibrutinib. A technique commonly used for this purpose is liquid chromatography-tandem mass spectrometry combined with isotope dilution in sample processing. Although this method provides a high degree of reliability, its use can be complicated with some specific factors and does not guarantee trouble-free analysis. This paper is focused on investigating issues related to the differential adsorption of ibrutinib and its D, D and C isotopically labeled analogues combined with instrument-specific carry-over. The results of the research point out the significantly different adsorption behavior of ibrutinib in fluidics of LC-MS compared with that of its D, D and C stable isotope labeled analogues, showing preferential adsorption of non-labeled compound. The investigation also pointed to a strong affinity of ibrutinib to polymeric surfaces under specific conditions, which has to be taken into consideration during sample preparation and analysis. Our work opens a new field for the discussion of scarcely reported problem related to the use of stable isotope labeled internal standards in LC-MS/MS analysis.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.jpba.2021.114366 | DOI Listing |