Fluorescence-encoded poly(methyl metharcylate) nanoparticles for a lateral flow assay detecting IgM autoantibodies in rheumatoid arthritis.

Anal Biochem

Brandenburg Technical University Cottbus - Senftenberg, Faculty of Natural Sciences, Universitätsplatz 1, D-01968, Senftenberg, Germany. Electronic address:

Published: November 2021

Rheumatoid arthritis (RA) belongs to the most often occurring autoimmune diseases in the world. For serological diagnosis, IgM auto-antibodies directed against the Fc portion of IgG referred to as rheumatoid factor are used as biomarkers. The autoantibody detection is usually done by ELISA. Such assays are reliable but are not suitable for point-of-care testing in contrast to lateral flow assays. Here, we report the development of a lateral flow assay based on carboxylated fluorescence-encoded poly(methyl methacrylate) nanoparticles. Poly(methyl methacrylate) is a non-toxic plastic with an excellent biocompatibility and high optical transparency which promises especially high sensitive fluorescence detection thereby leading to very sensitive assays. We could detect a positive signal in samples with a nephelometric reading down to 0.4 U/mL. By analyzing 30 sera of patients with a RA diagnosis and 34 sera of healthy test subjects we could confirm positive ELISA results in 72% of all cases and negative ELISA results in 97% of all cases.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ab.2021.114389DOI Listing

Publication Analysis

Top Keywords

lateral flow
12
fluorescence-encoded polymethyl
8
flow assay
8
rheumatoid arthritis
8
polymethyl methacrylate
8
polymethyl metharcylate
4
metharcylate nanoparticles
4
nanoparticles lateral
4
assay detecting
4
detecting igm
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!