Background:  Current near-infrared spectroscopy (NIRS)-based systems for continuous flap monitoring are limited to flaps which carry a cutaneous paddle. As such, this useful and reliable technology has not previously been applicable to muscle-only free flaps where other modalities with substantial limitations continue to be utilized.

Methods:  We present the first NIRS probe which allows continuous monitoring of local tissue oxygen saturation (StO) directly within the substance of muscle tissue. This probe is flexible, subcentimeter in scale, waterproof, biocompatible, and is fitted with resorbable barbs which facilitate temporary autostabilization followed by easy atraumatic removal. This novel device was compared with a ViOptix T.Ox monitor in a porcine rectus abdominus myocutaneous flap model of arterial and venous occlusions. During these experiments, the T.Ox device was affixed to the skin paddle, while the novel probe was within the muscle component of the same flap.

Results:  The intramuscular NIRS device and skin-mounted ViOptix T.Ox devices produced very similar StO tracings throughout the vascular clamping events, with obvious and parallel changes occurring upon vascular clamping and release. The normalized cross-correlation at zero lag describing correspondence between the novel intramuscular NIRS and T.Ox devices was >0.99.

Conclusion:  This novel intramuscular NIRS probe offers continuous monitoring of oxygen saturation within muscle flaps. This experiment demonstrates the potential suitability of this intramuscular NIRS probe for the task of muscle-only free flap monitoring, where NIRS has not previously been applicable. Testing in the clinical environment is necessary to assess durability and reliability.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9469635PMC
http://dx.doi.org/10.1055/s-0041-1732361DOI Listing

Publication Analysis

Top Keywords

intramuscular nirs
16
flap monitoring
12
nirs probe
12
near-infrared spectroscopy
8
muscle-only free
8
continuous monitoring
8
oxygen saturation
8
vioptix tox
8
tox devices
8
vascular clamping
8

Similar Publications

Power Doppler ultrasonography is used to measure blood flow within a given muscle, otherwise known as intramuscular blood flow. However, it is not fully understood how intramuscular blood flow and muscle oxygenation change with repetitive muscle contraction. The present study was conducted to assess changes in intramuscular blood flow and muscle oxygenation of the vastus lateralis (VL) during intermittent and incremental contractions.

View Article and Find Full Text PDF

Muscle microvascular oxygen delivery limitations during the contraction phase of intermittent maximal effort contractions.

Eur J Appl Physiol

September 2024

School of Kinesiology, Applied Health, and Recreation, Oklahoma State University, 180 Colvin Recreation Center, Stillwater, OK, 74078, USA.

Purpose: The end-test torque (ETT) during intermittent maximal effort contractions reflects the highest contraction intensity at which a muscle metabolic steady-state can be attained. This study determined if ETT is the highest intensity at which the contraction phase of intermittent exercise does not limit the matching of microvascular oxygen delivery to muscle oxygen demand.

Methods: Microvascular oxygenation characteristics of the biceps brachii muscle were measured in sixteen young, healthy individuals (8M/8F, 22 ± 3 years, 80.

View Article and Find Full Text PDF

Improving market (cull) dairy cows' carcass traits and meat quality.

J Dairy Sci

December 2024

Department of Population Medicine, University of Guelph, Guelph, ON N1G 2W1, Canada. Electronic address:

Despite a high volume of market (cull) dairy cows entering the food chain every year, beef from market dairy cows is largely considered insignificant when compared with beef from beef cattle in the market and is widely thought to be usable for ground beef only. Our study aimed to evaluate the effects of feeding dairy market cows on a lactating TMR diet before slaughter on carcass and beef quality traits. Forty-three Holstein market cows were randomly assigned into 2 treatments: fed or direct.

View Article and Find Full Text PDF

Despite the importance of intramuscular fat (IMF) to eating quality, as yet no methodology has been widely adopted by the whole of industry in Australia to measure it routinely. Thus, a study was conducted to investigate the potential for a Near Infra-Red (NIR) device to predict the IMF content of the loin from spectra collected on the topside which is externally located on a hanging carcase and therefore easily accessible. To this end, NIR spectra were collected from topsides (m.

View Article and Find Full Text PDF

Trimodal wireless intramuscular device detects muscle pressure, flow, and oxygenation changes in porcine model of lower extremity compartment syndrome.

Eur J Orthop Surg Traumatol

August 2024

Division of Plastic and Reconstructive Surgery, Department of Surgery, School of Medicine, Washington University, 660 S. Euclid Avenue, St. Louis, MO, 63110, USA.

Purpose: Compartment syndrome remains difficult to diagnose early in its clinical course. Pressure transducer catheters have been used to directly measure intracompartmental pressure (ICP), but this method is unreliable, with a false positive rate of 35%. We have previously used intramuscular near infrared spectroscopy to detect changes in tissue oxygen saturation (StO) in response to increasing ICP using a novel implantable probe.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!