Background: Spinal artery ischemia (SCI) events can result from over coverage of the descending thoracic aorta with a coated stent during Thoracic Endovascular Aortic Repair (TEVAR). The aim of this study was to determine whether a new distal perforating stent could reduce the incidence of spinal cord ischemia while remodeling the true lumen.
Methods: TBAD patients treated with Talos stent in the vascular surgery Department of Yan 'an Hospital affiliated to Kunming Medical University between December 2017 and October 2019 were retrospectively analyzed to investigate the short-term safety and effectiveness of Talos stent.
Results: A total of the 20 patients, including 14 males and 6 females, with an average age of 52.65 ± 8.98 years (range 37-68 years), were included in the analysis. Stent-grafts were successfully implanted in all patients under local anesthesia, with a technical success rate of 100%. The average operation time was 50.75 ± 13.01 min. A total of 2 cases (10%) presented chest pain associated with intercostal artery ischemia that was relieved on the 3rd and 5th postoperative day, respectively. Postoperative mean follow-up was 16.15 ± 3.99 months. No paraplegia or other complications occurred. And stenting did not induce new tears. No migration, deformation, or fracture of the stents occurred. There was a significant difference in the remolding of the true lumen preoperatively and at 12 months postoperatively (P < 0.05).
Conclusions: Talos stent has achieved satisfactory clinical treatment results in short term.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8456669 | PMC |
http://dx.doi.org/10.1186/s12872-021-02270-5 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!