Pulse oximetry is used widely to titrate oxygen therapy and for triage in patients who are critically ill. However, there are concerns regarding the accuracy of pulse oximetry in patients with COVID-19 pneumonitis and in patients who have a greater degree of skin pigmentation. We aimed to determine the impact of patient ethnicity on the accuracy of peripheral pulse oximetry in patients who were critically ill with COVID-19 pneumonitis by conducting a retrospective observational study comparing paired measurements of arterial oxygen saturation measured by co-oximetry on arterial blood gas analysis (SaO ) and the corresponding peripheral oxygenation saturation measured by pulse oximetry (S O ). Bias was calculated as the mean difference between SaO and S O measurements and limits of agreement were calculated as bias ±1.96 SD. Data from 194 patients (135 White ethnic origin, 34 Asian ethnic origin, 19 Black ethnic origin and 6 other ethnic origin) were analysed consisting of 6216 paired SaO and S O measurements. Bias (limits of agreement) between SaO and S O measurements was 0.05% (-2.21-2.30). Patient ethnicity did not alter this to a clinically significant degree: 0.28% (1.79-2.35), -0.33% (-2.47-2.35) and -0.75% (-3.47-1.97) for patients of White, Asian and Black ethnic origin, respectively. In patients with COVID-19 pneumonitis, S O measurements showed a level of agreement with SaO values that was in line with previous work, and this was not affected by patient ethnicity.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8653100PMC
http://dx.doi.org/10.1111/anae.15581DOI Listing

Publication Analysis

Top Keywords

pulse oximetry
20
ethnic origin
20
patient ethnicity
16
covid-19 pneumonitis
16
oximetry patients
12
patients covid-19
12
sao measurements
12
ethnicity accuracy
8
accuracy peripheral
8
peripheral pulse
8

Similar Publications

Brachycephalic breeds suffer from respiratory distress known as brachycephalic obstructive airway syndrome (BOAS) and the multiple comorbidities associated with it. Targeted breeding toward a more BOAS-free phenotype requires accurate and least invasive detection of BOAS severity grades that are accessible and accepted by the breeders and kennel clubs. This study aimed to compare the-outcome of morphometric anatomical examination with functional tests such as exercise tests and plethysmography for the detection of BOAS severity in a group of 84 French Bulldogs.

View Article and Find Full Text PDF

Heterogeneity of Intermediate Care Organization Within a Single Healthcare System.

Crit Care Explor

January 2025

Division of Pulmonary and Critical Care Medicine, Department of Medicine, Johns Hopkins University, Baltimore, MD.

Intermediate care (IC) is prevalent nationwide, but little is known about how to best organize this level of care. Using a 99-item cross-sectional survey assessing four domains (hospital and physical IC features, provider and nurse staffing, monitoring, and interventions/services), we describe the organizational heterogeneity of IC within a five-hospital healthcare system. Surveys were completed by nurse managers from 12 (86%) of 14 IC settings.

View Article and Find Full Text PDF

ACCREDIT: Validation of clinical score for progression of COVID-19 while hospitalized.

Glob Epidemiol

June 2025

Instituto Nacional de Infectologia Evandro Chagas, Fundação Oswaldo Cruz, Brazil.

Unlabelled: COVID-19 is no longer a global health emergency, but it remains challenging to predict its prognosis.

Objective: To develop and validate an instrument to predict COVID-19 progression for critically ill hospitalized patients in a Brazilian population.

Methodology: Observational study with retrospective follow-up.

View Article and Find Full Text PDF

Pulse oximetry at two sensor placement sites in conscious foals.

Acta Vet Scand

January 2025

Department of Equine and Small Animal Medicine, Faculty of Veterinary Medicine, University of Helsinki, Viikintie 49, 00014, Helsinki, Finland.

Background: Pulse oximetry has not been thoroughly evaluated for assessment of oxygenation in conscious foals. Compared with invasive arterial blood sampling, it is a painless and non-invasive method for real-time monitoring of blood oxygen saturation. The aim of this prospective clinical study was to evaluate the usability, validity, and reliability of pulse oximetry at two measuring sites (lip and caudal abdominal skin fold) for blood oxygen saturation measurement in conscious foals with and without respiratory compromise.

View Article and Find Full Text PDF

Background: This study aimed to enhance the scope of neonatal congenital heart disease (CHD) screening by evaluating the effectiveness of training personnel in CHD screening using the "dual-index" method, combining pulse oximetry with cardiac murmur auscultation.

Methods: From 2019 to 2022, a total of 2374 screening personnel from the Xinjiang, Yunnan, Hainan, Fujian, and Anhui provinces underwent training in neonatal CHD screening using the "dual-index" method, which involves pulse oximetry and cardiac murmur auscultation. Pre- and post-training assessments were conducted using a neonatal CHD screening knowledge questionnaire, distributed through the Questionnaire Star platform, to evaluate the impact of the training.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!