Medical linear-accelerator-based stereotactic radiosurgery (SRS) using a stereotactic apparatus or image-guided radiotherapy system for intracranial lesions is performed widely in clinical practice. In general, Winston-Lutz (WL) tests using films or electric portal imaging devices (EPIDs) have been performed as pre-treatment and routine quality assurance (QA) for the abovementioned treatment. Two-dimensional displacements between the radiation isocentre and mechanical isocentre are analysed from the test; therefore, it is difficult to identify the three-dimensional (3D) isocentre position intuitively. In this study, we developed an innovative 3D WL test for SRS-QA using a novel radiochromic gel dosimeter based on a polyvinyl alcohol-iodide (PVA-I) complex that can be reused after annealing. A WL gel phantom that was consisted of the PVA-I gel dosimeter poured into a tall acrylic container and an embedded small tungsten sphere was used as a position detector. A flatbed scanner was used to analyse the isocentre position. The measured 3D isocentre accuracy from the gel-based WL test was within 0.1 mm compared with that obtained from the EPID-based WL test. Furthermore, excellent reusability of the WL gel phantom was observed in long-term SRS isocentre verification, in which clinical SRS cases involving repeated irradiation and annealing were analysed. These results demonstrate the high accuracy and reliable evaluation of the isocentre position using an innovative test. In addition, the clinical-based routine SRS-QA using the PVA-I gel dosimeter demonstrates a highly convenience while affording an easy and fast analysis process.

Download full-text PDF

Source
http://dx.doi.org/10.1088/1361-6560/ac279dDOI Listing

Publication Analysis

Top Keywords

gel dosimeter
16
isocentre position
12
polyvinyl alcohol-iodide
8
alcohol-iodide pva-i
8
radiochromic gel
8
innovative test
8
gel phantom
8
pva-i gel
8
isocentre
7
test
6

Similar Publications

A gamma cell unit is used for gamma irradiation of small volume samples. Due to the sample volume, there is a dose distribution, which is imperative to be known for appropriate irradiation. In this study, samples of agarose Fricke gel dosimeters were prepared for the dose mapping of the gamma cell unit available in our laboratory.

View Article and Find Full Text PDF

Background And Purpose: During an end-to-end (E2E) test on the online workflow of the MR-linac, the performance of the treatment starting from the acquisition of pre-treatment MRI scans and ending with dose delivery is quantified. In such a test, the geometrical accuracy of the entire workflow is assessed. Ideally, the 3D geometrical accuracy of dose delivery on an MR-linac should be assessed using dosimeters that provide 3D dose distributions.

View Article and Find Full Text PDF

Background: Comprehensive quality assurance (QA) for a seamless workflow of high-dose-rate brachytherapy, from imaging to planning and irradiation, is uncommon, and QA of the source dwell position is performed in one- or two-dimensions. Gel dosimetry using magnetic resonance imaging (MRI) is effective in verifying the three-dimensional distribution of doses for image-guided brachytherapy (IGBT). However, MRI scanners are not readily accessible, and MRI scanning is time-consuming.

View Article and Find Full Text PDF

Quantitative spatial visualization of X-ray irradiation via redox reaction by dynamic nuclear polarization magnetic resonance imaging.

Free Radic Biol Med

November 2024

Department of Radiology, Gifu University, Gifu, Japan; Innovation Research Center for Quantum Medicine, Gifu University, Gifu, Japan; Department of Radiology, Frontier Science for Imaging, School of Medicine, Gifu University, Gifu, Japan.

The dose of X-ray irradiation is commonly measured by point assessment with an ionization chamber dosimeter. However, to achieve spatially accurate delivery of X-ray to avoid the exposure to normal tissues, an accurate imaging method for spatially and quantitatively detecting exposure is required. Herein, we present a novel method to visualize X-ray exposure using low-field dynamic nuclear polarization magnetic resonance imaging (DNP-MRI) with nitroxyl radical tempol as the chemical dosimeter.

View Article and Find Full Text PDF
Article Synopsis
  • This study focuses on using a surfactant hydrogel dosimeter (SHD) to accurately measure 3D absorbed dose distributions in carbon ion radiotherapy, which is important for targeting tumors effectively.
  • The researchers set up a cubic target region and conducted carbon ion beam irradiation, measuring the response of the SHD using optical computed tomography to gather data about the absorbed dose.
  • Results showed that the optical attenuation coefficients (OAC) measured were lower than the expected absorbed dose, indicating discrepancies that may affect precise dosimetry in treatment planning.
View Article and Find Full Text PDF