Prozone rates in the solid-phase platelet crossmatch assay and correlation with class I HLA antibody levels.

Transfusion

Division of Transfusion Medicine, Department of Laboratory Medicine & Pathology, Mayo Clinic, Rochester, Minnesota, USA.

Published: November 2021

Background: Solid-phase platelet crossmatch (PXM) testing is used to help manage patients with platelet transfusion-refractoriness. Recently, we published the first report of false-negative PXM results from prozone effect that was mitigated using sample dilution. This study aimed to describe the prevalence of PXM prozone effect and the levels of class I HLA antibodies (HLA-Abs) associated with positive PXM results and with false-negative PXM results from prozone effect.

Study Design And Methods: A cross-sectional study of patients undergoing PXM testing from July 2019 through December 2020 was performed. All PXM tests were run simultaneously using undiluted and 1:4 diluted patient plasma. Prozone effect was defined as a negative PXM result using undiluted patient plasma but a positive PXM result using 1:4 diluted patient plasma.

Results: Among 59 patients, 830 individual ABO-compatible PXM results yielded an overall positivity rate of 25.8% (214/830) and a false-negative rate from prozone effect of 4.7% (10/214). Among the 28 patients with class I HLA-Ab testing and no other anti-platelet antibodies, maximum HLA-Ab mean fluorescence intensity (MFI) was significantly associated with a positive PXM result (p < .0001; AUC approx. 0.9) and categorized into negative (<3700), indeterminate (3700-10300), and positive (>10300) maximum HLA-Ab MFI zones. Maximum HLA-Ab MFI, however, was not associated with prozone effect (p = .17; AUC approx. 0.6).

Discussion: While there is a strong predictive association between class I HLA-Ab levels and positive PXM results, PXM prozone effect is a common occurrence not associated with class I HLA-Ab levels, so additional testing with diluted samples should be considered.

Download full-text PDF

Source
http://dx.doi.org/10.1111/trf.16660DOI Listing

Publication Analysis

Top Keywords

pxm prozone
12
positive pxm
12
pxm result
12
pxm
11
solid-phase platelet
8
platelet crossmatch
8
class hla
8
pxm testing
8
false-negative pxm
8
associated positive
8

Similar Publications

Background: Platelet-transfusion refractory (PR) patients do not achieve expected post-transfusion platelet counts. We investigate suspected PR patients with post-transfusion platelet counts, indirect platelet antibody screens (ind-PAS), Class I HLA antibody tests (HLA-Scr), and physical platelet crossmatch (PXM) studies.

Study Design And Methods: The three following cases describe possible pitfalls of laboratory tests used in PR workup and management.

View Article and Find Full Text PDF

Prozone rates in the solid-phase platelet crossmatch assay and correlation with class I HLA antibody levels.

Transfusion

November 2021

Division of Transfusion Medicine, Department of Laboratory Medicine & Pathology, Mayo Clinic, Rochester, Minnesota, USA.

Background: Solid-phase platelet crossmatch (PXM) testing is used to help manage patients with platelet transfusion-refractoriness. Recently, we published the first report of false-negative PXM results from prozone effect that was mitigated using sample dilution. This study aimed to describe the prevalence of PXM prozone effect and the levels of class I HLA antibodies (HLA-Abs) associated with positive PXM results and with false-negative PXM results from prozone effect.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!