A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 176

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016

File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 316
Function: require_once

A Wi-Fi-Based Mask-Type Laryngoscope for Telediagnosis During the COVID-19 Pandemic: Instrument Validation Study. | LitMetric

A Wi-Fi-Based Mask-Type Laryngoscope for Telediagnosis During the COVID-19 Pandemic: Instrument Validation Study.

J Med Internet Res

Biomedical Engineering Research Center, Asan Institute for Life Sciences, Asan Medical Center, Seoul, Republic of Korea.

Published: October 2021

Background: Owing to the COVID-19 pandemic, social distancing has become mandatory. Wireless endoscopy in contactless examinations promises to protect health care workers and reduce viral spread.

Objective: This study aimed to introduce a contactless endoscopic diagnosis system using a wireless endoscope resembling a mask.

Methods: The Wi-Fi-based contactless mask endoscopy system comprises a disposable endoscope and a controller. First, the effective force applied by the tip during insertion was evaluated in a simple transoral model consisting of a force sensor on a simulated oropharynx wall. Second, the delay in video streaming was evaluated by comparing the frame rate and delays between a movement and its image over direct and Wi-Fi connections. Third, the system was applied to a detailed laryngopharyngeal tract phantom.

Results: The smartphone-controlled wireless endoscopy system was successfully evaluated. The mean, maximum, and minimum collision forces against the wall of the transoral model were 296 mN (30 gf), 363 mN (37 gf), and 235 mN (24 gf), respectively. The delay resulting from the wireless connection was 0.72 seconds. Using the phantom, an inexperienced user took around 1 minute to orient the endoscope to a desired area via the app.

Conclusions: Device articulation does not pose a significant risk of laryngopharyngeal wall penetration, and latency does not significantly impede its use. Contactless wireless video streaming was successful within the access point range regardless of the presence of walls. The mask endoscope can be controlled and articulated wirelessly, minimizing contact between patients and device operators. By minimizing contact, the device can protect health care workers from infectious viruses like the coronavirus.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8525626PMC
http://dx.doi.org/10.2196/31224DOI Listing

Publication Analysis

Top Keywords

covid-19 pandemic
8
wireless endoscopy
8
protect health
8
health care
8
care workers
8
endoscopy system
8
transoral model
8
video streaming
8
minimizing contact
8
wireless
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!