Dosimetric evaluation for temporomandibular joint cone beam computed tomography exams using different field of view.

Biomed Phys Eng Express

Division of Oral Radiology, Faculdade São Leopoldo Mandic, Instituto de Pesquisas São Leopoldo Mandic, Campinas, São Paulo (SP), 13045-755, Brazil.

Published: September 2021

To optimize the absorbed organ dose in relation to the field of view for temporomandibular joint examinations in four cone beam computed tomography devices.An anthropomorphic adult head and neck phantom, and 192 LiF dosimeters (TLD-100) were used. The dosimeters were placed in the region corresponding to the lens, parotid glands, submandibular glands, and thyroid. Small, medium and large FOVs were selected on Orthopantomograph OP300 Maxio, PaX-i3D Smart, ORTHOPHOS XG, and i-CAT Next Generation device when it was possible.A wide range of absorbed dose values was recorded for all organs due to the different exposure parameters of each device. The radiosensitive organ with the highest dose was the parotid glands. The devices with 5 × 5 cm FOV recorded a lower dose in this protocol, while for the device without a small FOV (≤5 × 5 cm), the lowest dose was observed with the large FOV (6 × 16 cm).We recommend a double exposure with an FOV of 5 × 5 cm in the OP300 Maxio, PaX-i3D Smart, and ORTHOPHOS XG device, while in the i-CAT Next Generation device, a single exposure FOV of 6 × 16 cm is indicated.

Download full-text PDF

Source
http://dx.doi.org/10.1088/2057-1976/ac240aDOI Listing

Publication Analysis

Top Keywords

temporomandibular joint
8
cone beam
8
beam computed
8
computed tomography
8
field view
8
parotid glands
8
op300 maxio
8
maxio pax-i3d
8
pax-i3d smart
8
smart orthophos
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!