Detecting Preanalytical Errors Using Quality Indicators in a Hematology Laboratory.

Qual Manag Health Care

Department of Medical Laboratory Science, College of Applied Medical Sciences, University of Hail, Saudi Arabia (Drs Alshaghdali, Rezgui, and JC Alcantara and Ms TY Alcantara); Department of Medical Laboratory Science, School of Pharmacy, College of Health Sciences, University of Wyoming, Casper (Dr Cruz); and Department of Clinical Laboratory, Maternity and Pediatric Hospital, Hail, Saudi Arabia (Messrs Alshammary and Almotairi).

Published: June 2022

Background And Objectives: Monitoring laboratory performance continuously is crucial for recognizing errors and fostering further improvements in laboratory medicine. This study aimed to review the quality indicators (QIs) and describe the laboratory errors in the preanalytical phase of hematology testing in a clinical laboratory.

Methods: All samples received in the Hematology Laboratory of the Maternity and Pediatric Hospital in Hail for 3 years were retrospectively reviewed and evaluated for preanalytical issues using a set of QIs. The rate of each QI was compared to the quality specifications cited in the literature.

Results: A total of 95002 blood samples were collected for analysis in the hematology laboratory from January 2017 through December 2019. Overall, 8852 (9.3%) were considered to show preanalytical errors. The most common were "clotted specimen" (3.6%) and "samples not received" (3.5%). Based on the quality specifications, the preanalytical QIs were classified generally as low and medium level of performance. In contrast, the sigma-based performance level indicates acceptable performance on all the key processes. Further analysis of the study showed a decreasing rate of preanalytical errors from 11.6% to 6.5%.

Conclusions: Preanalytical errors remain a challenge to hematology laboratories. The errors in this case were predominantly related to specimen collection procedures that compromised the specimen quality. Quality indicators are a valuable instrument in the preanalytical phase that allows an opportunity to improve and explore clinical laboratory process performance and progress. Continual monitoring and management of QI data are critical to ensure ongoing satisfactory performance and to enhance the quality in the preanalytical phase.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9208812PMC
http://dx.doi.org/10.1097/QMH.0000000000000343DOI Listing

Publication Analysis

Top Keywords

preanalytical errors
16
quality indicators
12
hematology laboratory
12
preanalytical phase
12
preanalytical
8
quality specifications
8
errors
7
quality
7
laboratory
7
performance
6

Similar Publications

Evaluation of pre-analytical factors impacting urine test strip and chemistry results.

Clin Chem Lab Med

January 2025

Department of Laboratory Medicine and Pathobiology, University of Toronto, Toronto, ON, Canada.

Objectives: Careful consideration of the pre-analytical process for urine examination is essential to avoid errors and support accurate results and decision-making. Our objective was to assess the impact of various pre-analytical factors on urine test strip and quantitative chemistry results, including stability, tube type, fill volume, and centrifugation.

Methods: Residual random urine specimens were identified.

View Article and Find Full Text PDF

Preanalytical Sedimentation Effects on Hematological Tests.

J Appl Lab Med

December 2024

HealthPartners Medical Laboratory, Pathology Lab, Bloomington, MN, United States.

Introduction: Specimens suspected of errors related to low hemoglobin or changes in hemoglobin beyond that of clinically explained variations during hospital stays are frequently redrawn under the auspices that they are contaminated. When lack of an indwelling IV eliminates contamination as a possibility, evaluation of the specimen between the time of collection and testing should occur.

Methods: As part of a quality improvement project, we investigated the impact of sedimentation on collected blood specimens not immediately transferred to their respective tubes from a syringe.

View Article and Find Full Text PDF

Objectives: Urinalysis is widely used and is also frequently requested in emergency situations for screening hypovolemia, urinary tract infections, diabetes, ketoacidosis and hematuria. Our aim was to evaluate the impact of reporting urinary sediment in emergency department specimens with the Sysmex UN system.

Methods: We evaluated urinalyses requested by the emergency department over a three-month period and examined red blood cell count interference, compared leukocyte esterase dipsticks to cytofluorimetric leukocyte count and nitrites to cytofluorimetric bacterial count.

View Article and Find Full Text PDF

Ceftriaxone, a widely used antibiotic, is one of the most common drugs to cause drug-induced immune hemolytic anemia. In this report, we describe the effect of ceftriaxone on red blood cell parameters (low red blood cell count, low hematocrit, and high erythrocyte index values) in two pediatric patients without clinical symptoms of hemolytic anemia. Although automated hematology analyzers have helped to detect incorrect results, a peripheral blood smear examination was necessary for recognizing the erythrocyte agglutinins caused by ceftriaxone.