Purpose: The purpose of this study is to evaluate biomechanical accuracy of 3D printed anatomical vessels using a material jetting printer (J750, Stratasys, Rehovot, Israel) by measuring distensibility via intravascular ultrasound.

Materials And Methods: The test samples are 3D printed tubes to simulate arterial vessels (aorta, carotid artery, and coronary artery). Each vessel type is defined by design geometry of the vessel inner diameter and wall thickness. Vessel inner diameters are aorta = 30mm, carotid = 7mm, and coronary = 3mm. Vessel wall thickness are aorta = 3mm, carotid = 1.5mm, and coronary = 1mm. Each vessel type was printed in 3 different material options. Material options are user-selected from the J750 printer software graphical user interface as blood vessel wall anatomy elements in 'compliant', 'slightly compliant', and 'rigid' options. Three replicates of each vessel type were printed in each of the three selected material options, for a total of 27 models. The vessels were connected to a flow loop system where pressure was monitored via a pressure wire and cross-sectional area was measured with intravascular ultrasound (IVUS). Distensibility was calculated by comparing the % difference in cross-sectional area vs. pulse pressure to clinical literature values. Target clinical ranges for normal and diseased population distensibility are 10.3-44 % for the aorta, 5.1-10.1 % for carotid artery, and 0.5-6 % for coronary artery.

Results: Aorta test vessels had the most clinically representative distensibility when printed in user-selected 'compliant' and 'slightly compliant' material. All aorta test vessels of 'compliant' material (n = 3) and 2 of 3 'slightly compliant' vessels evaluated were within target range. Carotid vessels were most clinically represented in distensibility when printed in 'compliant' and 'slightly compliant' material. For carotid test vessels, 2 of 3 'compliant' material samples and 1 of 3 'slightly compliant' material samples were within target range. Coronary arteries were most clinically represented in distensibility when printed in 'slightly compliant' and 'rigid' material. For coronary test vessels, 1 of 3 'slightly compliant' materials and 3 of 3 'rigid' material samples fell within target range.

Conclusions: This study suggests that advancements in materials and 3D printing technology introduced with the J750 Digital Anatomy 3D Printer can enable anatomical models with clinically relevant distensibility.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8414686PMC
http://dx.doi.org/10.1186/s41205-021-00114-8DOI Listing

Publication Analysis

Top Keywords

'slightly compliant'
28
test vessels
16
vessel type
12
material options
12
'compliant' 'slightly
12
distensibility printed
12
compliant' material
12
material samples
12
material
11
vessels
9

Similar Publications

A cross-sectional study on the assessment of adherence to cardiovascular medications in Sudan heart center.

PLoS One

January 2025

Ambulatory Healthcare Services, Academic Affairs, Abu Dhabi Health Services (SEHA), Abu Dhabi, UAE.

Background: Non-adherence to cardiovascular medications is a global problem with clinical, economic, and humanistic consequences. Investigation of this problem may open the road for proper management of cardiovascular diseases.

Objective: Our objectives were to assess the level of adherence to, and to examine factors influencing adherence to, cardiovascular medications in subjects visiting a heart center in Sudan.

View Article and Find Full Text PDF

Purpose: To investigate image quality and agreement of derived cardiac function parameters in a novel joint image reconstruction and segmentation approach based on disentangled representation learning, enabling real-time cardiac cine imaging during free-breathing.

Methods: A multi-tasking neural network architecture, incorporating disentangled representation learning, was trained using simulated examinations based on data from a public repository along with MR scans specifically acquired for model development. An exploratory feasibility study evaluated the method on undersampled real-time acquisitions using an in-house developed spiral bSSFP pulse sequence in eight healthy participants and five patients with intermittent atrial fibrillation.

View Article and Find Full Text PDF

Objective: Phyllodes tumor (PT) is a variant of fibroepithelial proliferations of the breast, histologically demonstrating a leaf-like pattern. The WHO has categorized PTs as benign, borderline, or malignant based on their histological characteristics. The objective of this paper is to assess the clinicopathological factors with malignancy in PT of the breast.

View Article and Find Full Text PDF

Water quality assessment of rooftop harvested rainwater across different roof types in a semi-arid region of South Africa.

Water Environ Res

January 2025

Water and Environmental Management Research Group, Faculty of Science, Engineering and Agriculture, University of Venda, Thohoyandou, South Africa.

Uneven distribution of precipitation and climate change have led to water shortages, adversely impacting numerous countries worldwide. Rooftop rainwater harvesting (RWH) has emerged as a crucial method for providing water for domestic uses. However, there are concerns about the quality of rainwater collected from roofs, as it may be contaminated with pollutants such as metals and microbiological pathogens.

View Article and Find Full Text PDF

Background: Splenic stiffness is a potential imaging marker of portal hypertension. Normative spleen stiffness values are needed to define diagnostic thresholds.

Objective: To report stiffness measurements of the spleen in healthy children undergoing liver magnetic resonance (MR) elastography across MRI vendors and field strengths.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!