Detection of opioid effect with pupillometry.

Auton Neurosci

Department of Anesthesia and Perioperative Care, University of California San Francisco (UCSF), San Francisco, CA, United States of America. Electronic address:

Published: November 2021

Background: Opioids produce pupillary constriction but their impact on pupillary unrest and the dynamic parameters of the pupillary light reflex have not been characterized. Given the increasing use of portable pupillometers for care of critically ill patients, it is important to distinguish between opioid effects on the pupil versus those that have been reported to arise from traumatic and ischemic brain insults. We undertook this study to determine which pupillary responses are most profoundly and consistently affected by a progressive infusion of remifentanil.

Methods: We studied the effect of remifentanil on the pupil using two portable infrared pupillometers in 18 volunteers. One pupillometer measured pupillary unrest in ambient light (PUAL) and the other pupillometer measured neurological pupillary index (NPi), constriction velocity (CV), pupil diameter (PD), latency, and % reflex (% reflex) following a transient light flash. Remifentanil was administered at predetermined weight-adjusted rates to raise opioid effect site concentration up to a range known to produce respiratory depression and oxyhemoglobin desaturation, based on a previously published pharmacokinetic model.

Results: PUAL was ablated by remifentanil, declining 94 ± 6% from baseline at the time of maximum drug effect. Other pupillary measurements decreased 50-65% from baseline. NPi was unchanged. At the time of oxyhemoglobin desaturation, deviations in PD, CV, and % reflex were widely scattered, whereas PUAL consistently approached zero.

Conclusion: PUAL is a highly specific indicator of central opioid effect. As a non-invasive measure, it may provide useful data to clinicians who prescribe opioids.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.autneu.2021.102869DOI Listing

Publication Analysis

Top Keywords

pupillary unrest
8
pupillometer measured
8
oxyhemoglobin desaturation
8
pupillary
7
detection opioid
4
opioid pupillometry
4
pupillometry background
4
background opioids
4
opioids produce
4
produce pupillary
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!