Purpose: To compare different intraocular pressure (IOP) readings in corneas with intrastromal corneal ring segments (ICRS) taken by three different tonometers; Goldmann applanation tonometry (GAT), air puff tonometer, and ocular response analyzer (ORA) corneal-compensated IOP (ORA-IOPcc) and determine the relation of these measurements to different corneal parameters taken by Pentacam.

Methods: An observational cross-sectional analytic study included patients who underwent ICRS keraring implantation at 3 months. In each eye, the two rings were placed using the femtosecond laser assisted technique 5.5 or 6 mm from the center. IOP was measured using three different tonometers; GAT, air puff tonometer, and ocular response analyzer (ORA) corneal-compensated IOP (ORA-IOPcc).

Results: Fifty eyes of 30 patients (20 males and 10 females) aged 27.56 ± 6.38 years were included. IOP measurements by GAT, air puff tonometer, and ORA-IOPcc were 13.28 ± 2.13 mmHg, 10.47 ± 2.55 mmHg, and 13.19 ± 2.78 mmHg, respectively. Comparisons between air puff and each of GAT and ORA-IOPcc were statistically highly significant (-value <0.001).

Conclusion: IOP measurements taken by air puff tonometer were significantly lower than those taken by GAT and ORA-IOPcc. These differences were not constant across the pressure range but increased as the pressure values determined using GAT and ORA increased. ORA-IOPcc and GAT showed similar readings. No correlation was found between any of the IOP readings taken by the three tonometers and the central corneal thickness.

Download full-text PDF

Source
http://dx.doi.org/10.1177/11206721211042546DOI Listing

Publication Analysis

Top Keywords

air puff
16
gat air
12
puff tonometer
12
intrastromal corneal
8
corneal ring
8
ring segments
8
three tonometers
8
tonometer ocular
8
ocular response
8
response analyzer
8

Similar Publications

High-resolution awake mouse fMRI at 14 tesla.

Elife

January 2025

Athinoula A. Martinos Center for Biomedical Imaging, Department of Radiology, Harvard Medical School, Massachusetts General Hospital, Charlestown, United States.

High-resolution awake mouse functional magnetic resonance imaging (fMRI) remains challenging despite extensive efforts to address motion-induced artifacts and stress. This study introduces an implantable radio frequency (RF) surface coil design that minimizes image distortion caused by the air/tissue interface of mouse brains while simultaneously serving as a headpost for fixation during scanning. Furthermore, this study provides a thorough acclimation method used to accustom animals to the MRI environment minimizing motion-induced artifacts.

View Article and Find Full Text PDF

Background: The 3xTg-AD transgenic mouse model of Alzheimer's disease (AD) is an important tool to investigate the relationship between development of pathological amyloid-β (Aβ) and tau, neuroinflammation, and cognitive impairments. Traditional behavioral tasks assessing aspects of learning and memory, such as mazes requiring spatial navigation, unfortunately suffer from several shortcomings, including the stress of human handling and not probing species-typical behavior. The automated IntelliCage system was developed to circumvent such issues by testing mice in a social environment while measuring multiple aspects of cognition.

View Article and Find Full Text PDF

Purpose: To assess the ability of the Dolphin air-pulse aesthesiometer to present multiple stimuli, which are separated temporally (in sequence) or spatially (simultaneously).

Methods: Two studies were performed to explore the cooling effects induced by double air-puff stimuli generated by a novel aesthesiometer composed of two micro-blower integrated units. The stimuli were delivered sequentially or simultaneously at the same or different spatial locations to an in vitro eye model monitored using thermography.

View Article and Find Full Text PDF

Nano to micron-size combustion particles in smokers' homes: Magnetic properties of tobacco and cigarette ashes.

Environ Pollut

December 2024

Instituto de Geociencias (IGc), Universidad Nacional Autónoma de México (UNAM), Boulevard Juriquilla No. 3001, 76230, Querétaro, Mexico.

Cigarette emission comprises high-temperature combustion processes producing diverse harmful compounds, gases, and particulate matter that are spread through the air and can be taken by smokers and passive smokers involuntarily. We determined the magnetic properties of tobacco cigarettes from six commercial brands. We also monitored the emission of particulate matter in some smokers' homes using low-cost sensors and the PM dataset was studied through a time series analysis.

View Article and Find Full Text PDF

Dopamine release in striatal striosome compartments in response to rewards and aversive outcomes during classical conditioning in mice.

Neurosci Res

November 2024

Oral Physiology, Department of Oral Functional Science, Faculty of Dental Medicine and Graduate School of Dental Medicine, Hokkaido University, Hokkaido, Japan.

The striatum consists of two anatomically and neurochemically distinct compartments, striosomes and the matrix, which receive dopaminergic inputs from the midbrain and exhibit distinct dopamine release dynamics in acute brain slices. Striosomes comprise approximately 15 % of the striatum by volume and are distributed mosaically. Therefore, it is difficult to selectively record dopamine dynamics in striosomes using traditional neurochemical measurements in behaving animals, and it is unclear whether distinct dynamics play a role in associative learning.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!