Simulating Radial Pressure Waveforms with a Mock Circulatory Flow Loop to Characterize Hemodynamic Monitoring Systems.

Cardiovasc Eng Technol

Center for Devices and Radiological Health, Office of Science and Engineering Laboratories, U.S. Food and Drug Administration, Bldg. 62 Rm 1129, 10903 New Hampshire Ave., Silver Spring, MD, 20993, USA.

Published: April 2022

Purpose: Mock circulatory loops (MCLs) can reproducibly generate physiologically relevant pressures and flows for cardiovascular device testing. These systems have been extensively used to characterize the performance of therapeutic cardiac devices, but historically MCLs have had limited use for assessing patient monitoring systems. Here, we adapted an MCL to include peripheral components and evaluated its utility for qualitative and quantitative benchtop testing of hemodynamic monitoring devices.

Methods: An MCL was designed to simulate three physiological hemodynamic states: normovolemia, cardiogenic shock, and hyperdynamic circulation. The system was assessed for stability in pressure and flow values over time, repeatability, waveform morphology, and systemic-peripheral pressure relationships.

Results: For each condition, cardiac output was controlled to the nearest 0.2 L/min, and flow rate and mean arterial pressure remained stable and repeatable over a 60-s period (n = 5, standard deviation of ± 0.1 L/min and ± 0.84 mmHg, respectively). Transfer function analyses showed that the systemic-peripheral relationships could be adequately manipulated. The results from this MCL were comparable to those from other published MCLs and computational simulations. However, resolving current limitations of the system would further improve its utility. Three pulse contour analysis algorithms were applied to the pressure and flow data from the MCL to demonstrate the potential role of MCLs in characterizing hemodynamic monitoring systems.

Conclusion: Overall, the development of robust analysis methods in conjunction with modified MCLs can expand device testing applications to hemodynamic monitoring systems. Properly validated MCLs can create a stable and reproducible environment for testing patient monitoring systems over their entire operating ranges prior to clinical use.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8885788PMC
http://dx.doi.org/10.1007/s13239-021-00575-2DOI Listing

Publication Analysis

Top Keywords

hemodynamic monitoring
16
monitoring systems
16
mock circulatory
8
device testing
8
patient monitoring
8
pressure flow
8
monitoring
6
mcls
6
pressure
5
hemodynamic
5

Similar Publications

Optimizing the life of vascular access during follow-up.

J Cardiovasc Surg (Torino)

February 2025

Department of Vascular Surgery, ASST Settelaghi Universitary Teaching Hospital, University of Insubria, Varese, Italy.

Optimizing the longevity of vascular access in hemodialysis patients remains a critical aspect of patient care, given the significant role of arteriovenous fistulas (AVFs) and arteriovenous grafts (AVGs) in enabling effective dialysis. Vascular access complications, such as stenosis, thrombosis, and cannulation-related damage, continue to challenge both the functionality and the sustainability of these access points. Recent advancements underscore the importance of a robust follow-up strategy, integrating clinical evaluations with diagnostic tools like color Doppler ultrasound (CDU) and emerging interventional approaches such as drug-coated balloon (DCB) angioplasty.

View Article and Find Full Text PDF

Background: Multiple intraoperative hemodynamic parameters are associated with an increased risk of early allograft dysfunction (EAD) following living donor liver transplantation (LDLT); however, there is significant center-to-center variability in terms of which parameters are used. We sought to determine which intraoperative hemodynamic parameters are most predictive of EAD following LDLT.

Methods: This is a systematic review following PRISMA guidelines (PROSPERO ID: CRD42023409711).

View Article and Find Full Text PDF

Current advances in neurocritical care.

J Intensive Med

January 2025

Department of Critical Care Medicine, Beijing Shijitan Hospital, Capital Medical University, Beijing, China.

This review summarizes the current research advances and guideline updates in neurocritical care. For the therapy of ischemic stroke, the extended treatment time window for thrombectomy and the emergence of novel thrombolytic agents and strategies have brought greater hope for patient recovery. Minimally invasive hematoma evacuation and goal-directed bundled management have shown clinical benefits in treating cerebral hemorrhage.

View Article and Find Full Text PDF

Monitoring of perioperative tissue perfusion and impact on patient outcomes.

J Cardiothorac Surg

January 2025

Department of Anesthesiology, Zhongda hospital, Southeast University, No. 87 Dingjiaqiao, Nanjing City, 210009, Jiangsu Province, China.

Monitoring perioperative tissue perfusion is crucial in clinical anesthesia to protect organs and ensure patient safety. Indicators like hemodynamic parameters, tissue metabolism, and microcirculation markers are used for assessment. Studies show intraoperative hypotension negatively impacts outcomes, though blood pressure alone may not reflect tissue perfusion accurately.

View Article and Find Full Text PDF

Background: The prone position is commonly used in surgical procedures and is known to cause significant hemodynamic changes, particularly a reduction in cardiac output. This study aimed to explore the relationship between intra-abdominal pressure (IAP), airway pressure, and cardiac output during spine surgeries under general anesthesia.

Methods: Twenty-five patients undergoing elective spine surgery in the prone position were enrolled in this prospective observational study.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!